- 159 -
19.1.3.
Printouts interval for continuous transmission
Parameter <P4.1.3. INTERVAL> enables setting frequency of printouts for
continuous transmission.
Frequency of printouts is set in seconds, wherein the accuracy is 0.1s.
The user may set any time value ranging from 1000 to 0.1 s.
The setting is valid for continuous transmission in calibration unit and in current
unit activated by means of the balance (see parameter P4.2.1.2), it is also valid
for continuous transmission activated by command sent from a computer (see
point 22 COMMUNICATION PROTOCOL).
19.1.4.
Cooperation with E2R
E2R is a control system, it controls all weighments performed on a balance.
Upon activation of this option, operations on some Databases are possible only
with the use of a computer (options are not accessible on a balance). In order to
start cooperation with E2R system set <YES> value of <E2R SYSTEM>
parameter.
Caution:
<E2R System> parameter activation may be carried out by the device
manufacturer or authorized persons, exclusively.
19.2.
Printer
Submenu <PRINTER> allows the user to choose a port and a device, to which
the data is to be sent upon pressing
button on balance. Content of data
being transferred is set in parameter <PRINTOUTS/PRINTOUT GLP>.
Procedure:
−
press
button
−
enter the menu <P4 DEVICES>
−
enter menu group<P4.2 PRINTER>
−
set balance parameters related to cooperation with the printer
−
<P4.2.1 PORT> to which the printer is to be connected