- 47 -
•
as the indication returns to zero and pictograms
→
0
←
and
are lit, press
button,
The balance returns to weighing with the accuracy of the I weighing range, and II weighing range
pictogram and marker of the last but one digit are blanked.
11.2.11. Under-Pan Weighing
In standard analytical and precision balances, loads can be weighed under a weighing pan. Such
means of operation requires placing a balance in an uplifted position. RADWAG offers a rack for
under-pan weighing. The rack is an optional equipment offered for PS series balances.
For under-pan weighing follow below procedure:
•
Remove plastic hole plug located in the balance base,
•
Under the plug you can find a special holder with an opening dedicated for suspending the load
(fixed permanently)
•
The opening is used for fixing a hook intended for either load or a dedicated weighing pan,
once depositing a load one can start weighing
•
Upon completing under-pan weighing, put the plastic plug back onto its place.
CAUTION!
The suspension for hook must not be turned, twisted or manipulated in any direction. Such
actions may cause damage to balance mechanism.
Mass of all additional elements of under pan weighing kit, like the hook, weighing pan, string, etc. should be
zeroed by pressing
or
.
11.3. COUNTING PARTS OF THE SAME MASS
The standard version of a balance is equipped with an option of counting small objects of the same
mass.
When the function is initiated for the first time, mass standard weight equals 0.0000g. If the mass
standard weight is determined and has been used in <PARTS COUNTING> mode, then the software
accepts the most recently used mass value as the mass standard weight.
11.3.1. PARTS COUNTING Mode Settings
The software allows entering settings of each working mode. Some settings are identical for all
working modes. They are described in section relating to <
WEIGHING
> mode.
This section covers only settings for <
PARTS
COUNTING
> mode.
Summary of Contents for PS 10100 R2