31
Fig. 25 Window with changed weighing unit
Options in case when the main unit is [kg]:
g (gram), kg (kilogram),
ct (carat), lb (pound), oz (ounce)*, N (Newton)*.
*)
–
unit disabled for verified scales.
Options
in
case
when
the
main
unit
is
[g]:
g
(gram),
kg (kilogram), lb (pound)*, ct (carat), oz (ounce)*, N (Newton)*.
*)
–
unit disabled for verified scales.
8. SCALE PARAMETERS
You can adjust the scale to ambient conditions (filter level) or to your own
needs (Autozero), additionally you can specify minimum mass value.
The parameters are to be found in
>
tab.
Weighing parameters:
Filter
Enter this parameter to adjust your weighing device to ambient conditions.
The higher filter level, the longer the indication takes to stabilise. Values:
None
,
Very fast
,
Fast
,
Average
,
Slow
.
Median filter
Median filter enables eliminating short impulse interferences (e.g. mechanical
shocks). Values:
None
- median filter disabled;
0.5
,
1
,
1.5
,
2
,
2.5
- median
filter enabled.
Autozeroing
Enter this parameter to enable automatic control and correction of zero
indication. There are, however, some cases when this function can be a
disturbing factor for the measuring process, e.g. very slow placing of a load
on the weighing pan (load adding, e.g. pouring, filling). In such a case, it is
recommended to disable the function. Values:
NO
- function disabled,
YES
- function enabled.