29
Load the weighing pan steadily, avoid
applying mechanical shocks.
Place the loads centrally on the
weighing pan (eccentricity errors are
specified by EN 45501 standard, point
3.6.2).
Do not apply concentrated force (total
load in one point).
Avoid side loading, in particular side
shocks.
7.2. Zeroing
In order to zero mass indication in
„MwManager”
weighing window (top right
corner), press
button or trigger zeroing function. Zero indication and
the following pictograms are displayed:
and
. Zeroing process is
an equivalent for determining a new zero point, recognized by the balance
as precise zero. The scale can be zeroed only when the indication is stable.
Zeroing can be performed using external button connected to the input with
zeroing function assigned.
Indication can be zeroed only within ±2% range
of the maximum capacity. If the
zeroed value is above ±2%
of the maximum capacity, then the software indicates
a respective Err2 message.