- 15 -
1.8. User Manual Significance
Even if you are experienced and have already worked with this type of weighing instrument,
you are still obliged to read this user manual carefully prior switching the device on. This
user manual contains crucial information for correct operation of the module.
As long as you follow the guidelines your balance operation is correct and reliable.
1.9. Operator Training
The weighing device must be utilized and supervised only by personnel who have been
trained to use such type of a weighing device.
2. TRANSPORT AND STORAGE
2.1. Delivery Checklist
Upon delivery it is necessary to check the package and the device, make sure that your
package bears no signs of damage. Otherwise contact the manufacturer's representative.
2.2. Packaging
Keep all package elements should your device be transported in the future. Remember that
only original packaging can be used for shipping purposes. Prior packing, uncouple any
cables, remove any separable components (weighing pan, shields, inserts). The device
components shall be packed into an original packaging, thus being protected against
potential damage during transportation.
3. UNPACKING AND INSTALLATION
3.1. Workstation
•
Operate the device in a workroom free of vibrations and shakes, where there are no
air drafts nor dust.
•
Ambient air temperature cannot exceed the range of: +10 °C ÷ +40 °C,
•
Make sure that the relative humidity is not higher than 80%,
•
Make sure that the temperature changes slowly in the course of operation.
•
Place the weighing instrument either on a robust-design table or on a wall bracket,
which is both distant from heat sources and insusceptible to vibrations.
•
Take special precaution while weighing magnetic objects, as part of the balance is a
strong magnet. Should such loads be weighed, use under-pan weighing option in
order to remove the weighed load from an area influenced by the magnet. The hook
for under-pan weighing is installed in the weighing instrument’s base. Ground the
module’s base should static electricity affect the weighing result. The grounding
screw is located at the back of the balance base.
3.2. Unpacking
Cut the adhesive tape. Take the device out of the packaging. Open the accessory
box, take the device components out of it.