![RADWAG MAS 1. Y User Manual Download Page 128](http://html1.mh-extra.com/html/radwag/mas-1-y/mas-1-y_user-manual_781002128.webp)
- 128 -
List of density reports data:
1.
Sample number
[field with a value, if in Density mode <YES> value is
set for <Request sample number> parameter]
2.
Start date
3.
End date
4.
Density
[calculated density of a sample, liquid or air]
5.
Density
[calculated volume of tested sample,
density
measurement of solid body
]
6.
Procedure
[depending on selected procedure, data
contained on a report may change]
7.
Operator
[name of an operator carrying out the procedure]
8.
Product
[name of tested sample]
9.
Standard liquid
[name of a liquid used for determining density
of
solid object
]
10. Standard liquid density
[
density determination of solid object
]
11. Temperature
[temperature of liquid used for carrying out a
measurement, density determination of solid
body]
12. Sinker volume
[
density determination of a liquid
]
13.
Mass of steel standard
[
density determination of air
]
14. Mass of aluminium standard
[
density determination of air
]
15. Density of steel standard
[
density determination of air
]
16. Density of aluminium standard
[
density determination of air
]
17. Weighing 1
[mass measurement of tested sample or a
sinker in the air, or mass of measured steel
standard
18. Weighing 2
[mass measurement of tested sample or a
sinker in liquid, or mass of measured aluminium
standard]
30.8. Control Reports
Each product control carried out using SQC or PGC mode is sent to a printer or saved to
<
PGC>
database. Each control saved to the database features specific number assigned
at the moment of control completion.
Control number format:
X / y y / M M / d d / H H / m m
, where:
X – control type, marked as follows:
U – control regulated by the act,
Z – control completed by an operator,
W – control regulated by an internal criteria for PGC or SQC mode
yy - year of control end,
MM - month of control end,
dd - day of control end,
HH - hour of control end,
mm - minute of control end,
You can preview data on each control.
Procedure:
•
Enter <
Databases
> submenu in accordance with section 30 of this user manual.
Summary of Contents for MAS 1. Y
Page 1: ...WEIGHING MODULES MAS Y MPS Y MAS 1 Y USER MANUAL IMMU 25 01 08 18 EN www radwag com...
Page 9: ...9 1 GENERAL INFORMATION 1 1 Dimensions MAS Module MPS Module...
Page 10: ...10 MAS 1 Module...
Page 200: ...200...
Page 201: ...201...