![RADWAG MAS 1. Y User Manual Download Page 120](http://html1.mh-extra.com/html/radwag/mas-1-y/mas-1-y_user-manual_781002120.webp)
- 120 -
It is impossible to determine <
T2-
> and <
T2+
> errors. Their values are doubled
values of errors <T1-> and <T1+>.
Procedure – setting control parameters:
Press <
SETUP
> button in <
Mass control
> mode.
Select
<
Databases
> and <
Products
> parameter.
Select product for which all values are to be selected. Set required parameters and
return to home screen.
•
Select controlled product from the products database
<
>
.
•
Set required parameters, press
button to start automatic mass control.
•
Automatic control is started. The weighing instrument is tared and the feeder starts
dispensing the first element onto the weighing pan.
•
With the first element on the weighing pan, the feeder pauses, the element is
weighed; the first stable measurement is carried out.
•
The measurement is saved, the weighing instrument is zeroed, the next element is
dispensed onto the weighing pan. The weighing process is carried out analogously.
•
The cycle is repeated until all pieces determined for <Batch quantity> process are
weighed.
•
The following information is displayed during control process:
- Reference sample quantity
- Product
- Average,
- Standard deviation
- T1- errors quantity
- T1+ errors quantity
•
Upon completed measuring process of a particular batch, a report is generated, saved
to balance's database and printed on a balance-connected printer.
CAUTION:
Each control process can be aborted if necessary, to abort a process press
button. The process is stopped and control cancelled (no report is generated).
29.3. Printouts
Printouts option enables you to determine content of a particular standard printout, and to
define non-standard printout.
Standard printout
Standard printout comprises four internal sections containing different variables.
Variable settings: YES – print; NO – do not print.
Settings for header, weighing, and footer printouts are specified in section 14.5. For settings
of mass control report read below.
CAUTION:
In order to specify which measurement data is to be printed on a report it is necessary to
modify settings for <Weighing printout template> parameter. To enable particular
measurement data printout set <YES> value for it.
Particular reports content: