- 110 -
28.8. Non-Destructive Average Tare Control Mode
Prior control you can determine average tare by weighing packages. Procedure:
Go to setting window of
PGC
mode.
Activate <
Average Tare Determining
> function by pressing grey field of the main display of
<
SETTINGS
> parameters.
The following window is displayed:
Where:
Product
- Product name
-
0.25T
condition value, given in [g].
- Average mass of a packaging, given in [g].
- Standard deviation.
- Characteristics of sample's negative
T1
errors
- Characteristics of sample's negative
2T1
errors
- Controlled packaging net weight
- Packaging tare
- Packaging control status
Load
empty
- Command for process in progress with specified number of all
packages to be weighed
-
Work pane in a form of graph.
Graph concerns control of mass only, while estimating average
tare value, the preview of packaging mass distribution is not
possible.
- Information on control
- Abort control process
CAUTION:
If a product is to be controlled by “Non-destructive average tare” mode, then
accordingly to
Regulation, standard deviation “S” for its packaging mass, based on at least 10
measurements, cannot be greater than 0.25 of maximum permissible negative error T, for
nominal packaging mass.