![RADWAG MAS 1. Y User Manual Download Page 103](http://html1.mh-extra.com/html/radwag/mas-1-y/mas-1-y_user-manual_781002103.webp)
- 103 -
26.4. SQC Control Report
Example of SQC control report:
----- Report No.: W/16/07/15/07/45 -----
----------------------------------------
Balance type XA 4Y
Range 220 g
Reading unit 0.0001 g
Balance S/N 442566
Start date 16.07.15 07:41:55
End date 16.07.15 07:45:25
Operator Kowalski
Product TEST 01
Batch number
Nominal mass 50 g
Tare 0 g
[T1-] error value 1 g
[T2-] error value 3 g
[T1+] error value 1.5 g
[T2+] error value 4 g
Batch quantity 100
Measurements quantity 30
[T1-] errors quantity 0
[T2-] errors quantity 0
[T1+] errors quantity 0
[T2+] errors quantity 0
Min 50.0525 g
Max 50.3638 g
Average 50.291163 g
Sum 1508.7349 g
Standard deviation 0.133916 g
Method SQC
Result Positive
Measurements
1. 50.0525 g
2. 50.0525 g
3. 50.0525 g
4. 50.0525 g
----------------------------------------
Signature
........................................
Report template
Open <
Printouts/
SQC report printout template
> section to edit a template of a
report. Only variables with <YES> attribute are printed on a report.
27. PEAK HOLD
Peak Hold
is a working mode allowing you to snap value of maximum force applied to the
weighing pan during one loading.
Apart from standard settings for this mode (described in the
Weighing
mode section),
additional parameter has been introduced, it sets threshold value determining function
activation.
27.1. Mode Related Settings
The supplementary settings enable you to adjust the working mode to your needs and
requirements. To access the settings follow the below procedure.
Procedure:
1. Press grey workspace area.
2. The following submenus are displayed: Settings, Buttons, Information,
Printouts, Series, Profile.
3. Select <Settings> menu.
4. Functions related to Peak Hold mode are displayed.
THRESHOLD
– value determining start point for peak hold control performed by the
balance software. Remember to set this threshold prior to the measuring process.