18.4.11. Switch off continuous transmission in current unit
Syntax:
CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
18.4.12. Send all implemented commands
Syntax:
PC CR LF
Possible answers:
PC_- >_Z,T, TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC
– command executed,
the indicator have sent all the implemented commands.
18.5. Manual printouts
Users can general manual or automatic printouts from the scale. Manual
printouts can be performed after loading the pan and stabilizing indication
by pressing
.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1
2
3
4 -12
13
14
15
16
17
18
stability
space
sign
mass
space
unit
CR
LF