HX5.EX

4-Load-Cell Scales with EX Frame

FRAME ASSEMBLY MANUAL 

ITKM-003-01-0

8

-19-EN 

Summary of Contents for HX5.EX

Page 1: ...HX5 EX 4 Load Cell Scales with EX Frame FRAME ASSEMBLY MANUAL ITKM 003 01 08 19 EN...

Page 2: ...2 AUGUST 2019...

Page 3: ...to stop using the device immediately The damaged component must be replaced or repaired by RADWAG service immediately While installing the device it is necessary to follow strictly this service manual...

Page 4: ...4 CONTENTS 1 INTENDED USE 5 2 FRAME ASSEMBLY 6 3 WARRANTY CONDITIONS 8 4 SECURITY REQUIREMENTS 9...

Page 5: ...as explosion group IIB and IIA and as temperature class T1 T2 T3 T4 frames made of stainless steel can be used in zones 1 and 2 where there is a risk of explosion due to mixture of air with vapour mis...

Page 6: ...inimum Prepare recess for the frame The recess must be compliant with the requirements of zones 1 and 2 of hazardous area If there is a risk of an explosive atmosphere occurrence due to mixture of air...

Page 7: ...zardous area recess for the scale In order to prevent electrostatic charges accumulation it is necessary to connect the frame the platform and the indicator to the same equalizing circuit Spots marked...

Page 8: ...epresentatives C RADWAG does not bear any responsibility for damage or losses resulting from misuse or unauthorized use or servicing D The warranty does not cover mechanical damage caused by product e...

Page 9: ...on in dust atmospheres Diagram with areas of operation is presented in the first section of this service manual Explosion safety of ramps is ensured by Making frame mechanical design in accordance wit...

Page 10: ...10...