- 54 -
First declare the mass of a weight that is to be used for adjustment. The mass must
be
≥
30% Max capacity.
Once the mass of the weight is entered and confirmed, a message prompting the
user to remove the weight from the pan is displayed:
<REMOVE MASS>
(the
weighing pan must be empty). Follow the instruction, next press
button, the
scale determines weight of an unloaded pan, message <
CALIBRATION
> is shown
in the bottom line. Next, message <
PUT WEIGHT
> and mass value to be placed on
the weighing pan are displayed,
e.g.
200.000g
(depending on the scale type). Put
an external adjustment weight of displayed weight value on a pan and press
button
.
The scale determines the mass, message <
CALIBRATION
> is displayed in
the bottom line. On completing adjustment process the scale returns to submenu
P1.2 EXT.CALIB.
13.4
Report on adjustment – printout
At the end of each adjustment process or adjustment test, an adjustment report is
generated automatically and sent to communication port COM 1. The content of the
report is declared in menu P5.1 CAL REPORT.
The instruction on how to declare settings for this option is described in section
about printouts.
The report can be printed out via a printer connected to the device or it can be sent
to the computer and saved as a file for archiving purposes.