- 51 -
13.2
Adjustment menu parameters
P1
ADJUSTMENT
P1.1 INT.CALIB
|
[internal adjustment]
P1.2 EXT.CALIB
|
[external adjustment]
P1.3 USER
CALIBRATION
|
[user adjustment]
P1.4 CALIBRATION
TEST
|
[adjustment test]
P1.5 AUTO.CALIB.
| NONE/TIME/TEMP/BOTH [automatic
adjustment]
P1.6 AUTO.CALIB.C
|
[time of automatic
adjustment]
INT.CALIB – internal adjustment
Start of an internal adjustment process. Adjustment is carried out automatically, no
user assistance is required. If a weighing pan is loaded, the terminal displays
a command ordering to unload it. If weighing pan is loaded with a relatively small
weight (up to ±10% of max capacity) then adjustment process is carried out
automatically.
or
EXT.CALIB – external adjustment (by means of external adjustment weight)
Adjustment performed with an external weight, value of which is saved in factory
parameters.
USER CALIBRATION - user adjustment (by means of external adjustment weight)
Adjustment performed with an external weight of any mass within scale range,
however not lower than 30% of Max range.