- 30 -
•
Use arrow buttons to select the tare which is to be recalled and press
button,
•
The scale returns to the weighing mode, the display indicates entered tare
value with minus „–" sign.
or
•
For any operating mode, press
button.
•
Enter b3 <TARE> option.
•
Wait for the first packaging weight, recorded in tare database, to be
displayed.
•
Use arrow buttons to select the tare which is to be recalled and press
button.
•
The scale returns to the weighing mode, the display indicates entered tare
value with minus „–" sign.
AUTOTARE
Function allowing for automatic tarring of the packaging during the weighing
process when packaging mass for each of the product is different. Read section 9.5
for AUTOTARE function overview.
Deleting tare
Entered tare value can be deleted by pressing
button of the scale’s overlay or
by entering tare value of 0.000g (see description above).
9.4 WEIGHING mode settings
The software allows setup of operating parameters (filters, value release and
autozero option, deleting the last digit and other settings) separately for each
working mode.
It enables customizing the instrument and utilizing its properties depending on the
user’s needs and expectations, or on specific requirements for selected working
mode (e.g. DOSING); as a result the device operation is quick and easy.
Settings:
P2.2.1 – READOUT
P2.2.1.1 – FILTER [V.FAST/ FAST / NORMAL / SLOW / V. SLOW]
P2.2.1.2 – RESULT CONF. [REL. / FAST / FAST+REL.]
P2.2.1.3 – AUTOZERO [YES / NO]
P2.2.1.4 – LAST DIGIT [ALWAYS / NEVER / WHEN STAB.]
P2.2.1.5 – AMBIENT CONDITIONS (STABLE/UNSTABLE)
P2.2.2 – AUTOTARE [YES / NO]