Summary of Contents for APC series

Page 1: ...n balances APP C series M A N U F A C T U R E R O F E L E C T R O N I C W E I G H I N G I N S T R U M E N T S RADWAG 26 600 Radom Bracka 28 Street POLAND phone 48 48 38 48 800 phone fax 48 48 385 00 1...

Page 2: ...2 JUNE 2008...

Page 3: ...7 Connection of additional equipment 10 5 BASIC FUNCTIONS OF THE INSTRUMENT 11 6 KEYBOARD 12 7 USER MENU 13 7 1 Moving in user menu 15 7 1 1 Return to weighing function 15 8 WEIGHING 15 8 1 Choice of...

Page 4: ...emory by user 51 13 4 Defining thickness of solids and liquids 52 13 4 1 Testing thickness solids 52 13 4 2 Testing thickness of liquid 53 13 5 Summing function 53 14 FUNCTIONS FOR USING RS 232 55 14...

Page 5: ...5 19 ADDITIONAL EQUIPMENT 68 19 1 Vibration damping table 68 19 2 Additional display 68 19 3 Computer software 68...

Page 6: ...2 APP 30 C 2 Max capacity 8 kg 10 kg 25 kg 30 kg Tare range 8 kg 10 kg 25 kg 30 kg Min capacity 0 2 g 2 g Readability 0 01 g 0 1 g Repeatability 0 01 g 0 1 g Linearity 0 02 g 0 2 g Working temperatur...

Page 7: ...result Do not place any magnetic materials on the weighing pan as this can cause damage of the measuring system of the instrument Be sure to avoid impact shock and overloading the balance in excess o...

Page 8: ...of the original packaging in case it should be necessary to transport or return items at any time Only the original packaging should be used for return consignments Before dispatch disconnect all atta...

Page 9: ...container In case of static discharges may occur please ground the device Grounding bolt is placed at the back side of the balance s casing 4 2 Unpacking Cut safety tape off Take balance off the box T...

Page 10: ...ower adapter specified on the data plate of the power adapter should be compatible to the power from mains Please plug the adapter the socket at the back of the balance the display with light on and s...

Page 11: ...evel 9 Functions for calibration and weighing modes Counting pieces PIECES Control HiLo Control of deviations according to standard mass Perc A Perc B Thickness of liquids d_Li or solids d_Co Summing...

Page 12: ...lects the work mode MODE key Changes measure units UNITS key Sends information to external instrument PRINT or confirms parameter value or function ENTER PRINT ENTER Key Sets indication to zero Esc TA...

Page 13: ...t YES no printout of measurement date P2 5 PUSr YES no printout of user name P2 6 PPrJ YES no printout of project name P2 7 PId YES no printout of factory number of balance P2 8 PFr YES no printout of...

Page 14: ...c Functions P6 1 FFun choice of balance function P6 2 PcS YES no counting pieces P6 3 HiLo YES no check weighing P6 4 PrcA YES no deviation with reference to weighed standard mass P6 5 Prcb YES no dev...

Page 15: ...ll go back to weighing 8 WEIGHING Before weighing or if the conditions of work changes e g if temperature of surrounding is higher then 3 o C balance should be calibrated Balance should be loaded few...

Page 16: ...key If user presses key ON OFF once again balance is ready to following measurements 8 1 Choice of basic weighing unit Function is used to setting the initial unit with which the balance will start a...

Page 17: ...nt 7 1 2 return to weighing ATTENTION Balance after switching on will start with set basic unit 8 2 Choice of weighing unit temporary Function enables choice of unit with which mass on the pan will be...

Page 18: ...e to be accessible for user after pressing SETUP key Enter group P6 Func 8 3 1 Choice of modes quantity accessible for user Function enables user to set if after pressing RIGHT ARROW key all operating...

Page 19: ...y Balance will go back to displaying name ofsubmenu P6 1 FFun If function other than ALL is chosen procedure of returning to weighing differs from remaining ones If e g function PcS is chosen procedur...

Page 20: ...0 Press key Esc TARA balance will return to weighing 8 3 2 Setting contents of P6 1 FFun ALL parameters Function enables switching off modes not used by operator This causes faster access to used mode...

Page 21: ...weighing 9 MAIN SETTING PARAMETERS User can adjust balance to external conditions filter range or own needs autozero working memory of tare value These parameters are in group P3 rEAd These functions...

Page 22: ...edure using NAVIGATION BAR choose value of filter which is needed 1 filtr FASt fast 2 filtr StAnd standard 3 filtr Slouu slow Return to weighing see 7 1 2 return to weighing ATTENTION The higher filte...

Page 23: ...s it is advisable to adjust the balance by selecting method of result confirmation as FAST_rEL Fast or rEL Depending on selected option time of weighing will be shorter or longer Fast_rEL fast release...

Page 24: ...value smaller than declared AUTOZERO range e g one division balance will zero automatically and stable result marker and zero indication will be displayed When AUTOZERO function is on each result sta...

Page 25: ...25 9 4 Light off the last display digit Function enables light off the last digit on the display ALuuAYS always nEuEr never uu_StAb when is stable Return to weighing see 7 1 2 return to weighing...

Page 26: ...according to point 6 1 10 1 Backlight of weighing result bl OFF backlight off bl On backlight on bl Aut If result of weighing doesn t change for 10 seconds backlight will turn off automatically Retur...

Page 27: ...g key off bEEP On signal of pressing key on Return to weighing see 7 1 2 return to weighing 10 3 Printout of all balance parameters through RS 232 port After pressing F key balance parameters will be...

Page 28: ...is the balance calibration Calibration should be performed Before the weighing When long breaks are between following measure series When temperature inside the balance changes more than 3 C Kind of...

Page 29: ...ossibility of finishing series of measurements in 5 minutes balance will indicate CAL_30 message again process of calibration can be postponed frequently but fact that long postponing of calibration c...

Page 30: ...ay will indicate message about necessity of removing the load or CAL button P1 2 ECAL External calibration calibration with external weight which value is saved in factory menu of balance function una...

Page 31: ...actor which should decide about starting the automatic internal calibration nonE none of factors will cause start of calibration tEmP calibration with regard to change of temperature timE calibration...

Page 32: ...ermining the time after which automatic calibration will start P1 6 CALr printout from calibration report Setting printout of report after finished calibration no printout of report off YES printout o...

Page 33: ...result if balance is connected to computer or to printer through RS 232 joint printout of calibration test will take place Press ESC TARA key to return to previous screen Procedure Return to weighing...

Page 34: ...alance memory during the calibration Process proceeds automatically and display indicates its result if balance is connected to computer or to printer through RS 232 joint printout of calibration test...

Page 35: ...not load the pan 3 After this process the balance records results of the calibration in the memory and returns to weighing mode ATTENTION Pressing the ESC key stops the calibration process If during t...

Page 36: ...calibration the balance returns to submenu P1 2 uCAL 6 Return to weighing as in the point 5 1 2 If balance is verified user does not have the possibility to perform external calibration process 11 5 C...

Page 37: ...ce between calibration mass remembered by balance after last calibration and calibration mass determined during actual calibration and other information 12 SETTING PRINTOUTS FOR GLP PROCEDURES P2 GLP...

Page 38: ...do not print during report yes print during report 12 1 Setting the contents of printouts for GLP procedures P2 GLP group of parameters enabling to declare variables which will appear on calibration p...

Page 39: ...introduce name of user who operates the scale User has possibility to introduce the name consisting of max 8 alphanumeric characters Introduction the name is possible by the means of scale keyboard wi...

Page 40: ...r equivalents displayed by scale are shown on the drawing below Exemplary user name introduced to scale with capital letters OP1_WILK operator 1 WILK Exemplary user name introduced to scale with small...

Page 41: ...abling to print time of performed measurement P2 4 PdAt Option enabling to print data of performed measurement P2 5 PUSr Option enabling to print user name P2 6 PPrJ Option enabling to print project n...

Page 42: ...ANCE OPERATING MODES Weighing mode basic Counting pieces control referring mass of set standard Control of deviation referring mass of standard Definition thickness of solids and liquids Summing Press...

Page 43: ...functions is described in further part of this manual 13 1 Counting details of the same mass Balance in standard execution is equipped with option of counting small pieces of the same mass Counting p...

Page 44: ...if you want to choose one them e g 20pcs press ENTER PRINT and proceed as shown on picture If option FrEE was chosen you must enter quantity of sample which will serve to determine mass of single det...

Page 45: ...e of digit Confirm entered value by pressing ENTER PRINT key Display will indicate LoAd message place on the pan as many details as entered during option Press ENTER PRINT key balance will indicate qu...

Page 46: ...weighing see 7 1 2 return to weighing ATTENTION If user presses ENTER key when details are not placed on the pan message Er8 outr will be indicated for few second and balance will automatically retur...

Page 47: ...47 13 2 control referring mass of set standard Procedure of operating Enter to function During setting threshold values following dependences take place...

Page 48: ...hoice of set digit TOP DOWN ARROW choice of value of digit ENTER PRINT confirmation of entered values SET UPPER THRESHOLD RIGHT ARROW choice of set digit TOP DOWN ARROW choice of value of digit ENTER...

Page 49: ...n to weighing Return to weighing see 7 1 2 return to weighing 13 3 Control of deviation referring mass of standard Balance software enables control of deviation in of weighed loads mass referring mass...

Page 50: ...he pan load which mass will be accepted as standard press ENTER PRINT to confirm this operating mode after few seconds indication 100 00 will be displayed From this moment display will not indicate ma...

Page 51: ...ance memory by user Procedure Enter to function Display will show indication as above Using keys choice of set digit and choice of digit value set value of mass of standard enter it to balance memory...

Page 52: ...WAG service can render special software to determine thickness by balance accessible 13 4 1 Testing thickness solids Thickness of solids can be tested in one of 3 liquids H2O distilled water C2H5OH sp...

Page 53: ...and in tested liquid Result of testing is shown on display automatically after introducing mass of plunger 13 5 Summing function In sum procedure following components are added in order to make mixtur...

Page 54: ...54...

Page 55: ...g 14 FUNCTIONS FOR USING RS 232 Choose speed of transmission Define sent data as stable not stable Continuous work Setting data as additional element of parameter in printouts Turn on off sign of last...

Page 56: ...56 14 1 Speed of transmission Method of setting speed of transmission choose required speed of transmission 2400 bit s 4800 bit s 9600 bit s 19200 bit s Return to weighing see 7 1 2 return to weighing...

Page 57: ...Continuous transmission 14 2 1 Continuous transmission off in actual interval Procedure CntA no continuous transmission off CntA YES continuous transmission on Return to weighing see 7 1 2 return to...

Page 58: ...58 14 2 2 Continuous transmission of in basic units Procedure Cntb no continuous transmission off Cntb YES continuous transmission on Return to weighing see 7 1 2 return to weighing...

Page 59: ...to weighing Automatic operating takes place according to following scheme press TARE key to zero the balance display will indicate marker of stable measurement and marker of zero place the load balanc...

Page 60: ...data type sent through RS 232 joint only for non verified balances Procedure PStb no sending stable or temporary result of weighing PStb YES sending stable result of weighing Return to weighing see 7...

Page 61: ...functions Software enables setting operating of function automatic operating for automatic operating result will not be sent to computer or printer until indication of mass goes below set Lo net value...

Page 62: ...ts 8 Stop bit 1 Parity control none Sending weighing results to computer can take place manually after pressing PRINT key continuously after function activating or sending steering commend automatical...

Page 63: ...63 15 1 Schedules of connecting conductors WEIGHT COMPUTER 2 RxD 3 TxD 3 TxD 2 RxD 4 DTR 6 DSR 5 GND 5 GND 6 DSR 6 DTR 7 RTS 8 CTS 8 CTS 7 RTS...

Page 64: ...t is possible to do Example of printout with statistics from series of measurement 1 9 02 15 7 0016 g 2 9 02 39 5 0152 g 3 9 02 58 12 0171 g 4 9 03 15 9 9937 g 5 9 03 34 12 0169 g 6 9 03 48 22 0111 g...

Page 65: ...an error of exceeding the range on Mark space for positive values or for negative values mass 9 marks alignment to the right units 3 marks alignment to the left 15 4 2 Format of sent data for orders...

Page 66: ...unit 3 marks alignment to the left 16 WEIGHING LOADS UNDER BALANCE In standard analytical and precision balances can weigh load on suspension To use this function Remove plastic plug in basic of balan...

Page 67: ...rk mode on manual printout Function CONSTANCE WORK equivalent for key conb on Format CU 1 CR LF start continuous transmission off in actual interval Function MANUAL WORK equivalent for order conb OFF...

Page 68: ...Er10 cal calibration error change of mass or incorrect mass of standard mass 19 ADDITIONAL EQUIPMENT 19 1 Vibration damping table Thanks to this stable base all vibrations are eliminated Inside of tab...

Page 69: ...69 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 600 Radom Bracka 28 Street POLAND Central phone 48 48 38 48 800 phone fax 48 48 385 00 10 Sales Department 48 48 366 80 06 www radwag com...