ABL800 FLEX Operator's Manual
7. Replacements
Decontamination and Protein Removal programs
Decontamina-
tion program
The Decontamination program disinfects the liquid transport system and items in
direct contact with blood such as electrodes, pump tubing, etc.
CAUTION – Replacements before decontaminations
Do not perform the Decontamination Program without first replacing the
c
Cl
–
,
c
Glu,
c
Lac,
c
Crea A and
c
Crea B electrodes with dummy electrodes,
unless the membranes are to be replaced at this time. The Hypochlorite
Solution used in decontamination will damage the membranes.
It is recommended to perform Decontamination program once a month.
To perform Decontamination, do the following:
Step Action
Replace the
c
Cl
–
,
c
Glu,
c
Lac,
c
Crea A and
c
Crea B electrodes with
dummy electrodes, code 945-626.
1.
2.
Fill a syringe with 0.5 mL of S5362 Hypochlorite Solution.
Press
Menu
–
Start Programs
–
Auxiliary Programs –
Decontamination
.
(To cancel the program at this time, press
Close.
)
3.
4.
Open the syringe inlet flap and place the syringe tip in the inlet.
Press
Start
to start aspiration of the solution and the program to begin.
5.
6.
Remove the syringe and close the inlet flap as prompted on the screen.
When the program is completed, the analyzer returns to the
Ready
screen.
7.
Replace the dummy electrodes with the
c
Glu,
c
Lac,
c
Cl
–
,
c
Crea A
and
c
Crea B electrodes.
8.
9.
Perform a 2-point calibration.
Protein Removal
program
The Protein Removal program is a shorter version of the Decontamination program
and is used when required to clean the pH/BG and oximetry modules, pump
tubings and the inlets. Replacement of the
c
Glu,
c
Lac,
c
Crea and
c
Cl
–
electrodes
with dummy electrodes is not required.
The Protein Removal program must be performed once a week in order to maintain
the performance of the reference electrode membrane.
To perform Protein Removal, do the following:
Step Action
1.
Fill a syringe with 0.5 mL of S5362 Hypochlorite Solution.
Press
Menu
–
Start Programs
–
Auxiliary Programs – Protein
Removal
.
(To cancel the program at this time, press
Close
.)
2.
Continued on next page
7-21
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...