7. Replacements
ABL800 FLEX Operator's Manual
Replacing inlet gasket unit and inlet probe
Replacing the
inlet gasket unit
To replace the inlet gasket, do the following:
Inlet gasket unit
Mount
Step Action
1.
Remove the capillary and then the syringe inlet flaps and record the
action in the Hold mode.
2.
Remove the old inlet gasket unit by grabbing it on either side and
lifting it upward.
3.
Slide the new inlet gasket unit onto the mount and press down the top
of the unit to secure it in place.
Mount the syringe inlet flap and then the capillary inlet flap.
4.
Press
Restart
.
To replace the inlet probe, do the following:
Replacing the
inlet probe
Attachment
and clip
Tube
Inlet
probe
Step Action
1.
Remove the inlet flaps to enter the Hold mode.
2.
Remove the gasket unit – see
Replacing the inlet gasket unit
above.
3.
Lift the clip to open it.
4.
Lift the inlet probe out of the attachment.
Continued on next page
7-12
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...