7. Replacements
ABL800 FLEX Operator's Manual
General information
Viewing
replacement
requirements
To view the replacements that are due, do the following:
Step Action
1.
Check the following on the
Analyzer Status
screen:
•
Replacement schedule for the electrode membranes and the
electrodes
•
Messages referring to the membranes/electrodes or user activities
•
Traffic light color (G
REEN
= No replacements are due at the present
time; Y
ELLOW
= A replacement is due)
2.
Highlight the item to enter the screen for replacements of
electrodes/membranes – see the next page.
3.
Press
Log Activity
to enter the User Activity program to view user
activities list – see
chapter 3
in this manual.
4.
Press
Close
to exit to the
Ready
screen.
The Hold
mode ensures that all wet section activities are suspended and that the
proper restart sequence will follow the exit from the Hold mode.
Entering the
Hold mode
To enter the Hold mode, do one of the following:
•
Go to
Menu
–
Analyzer Status
–
Electrodes and Other
. Press
Replace
.
•
Remove the cover to the measuring modules.
•
Open the window to the electrode module and wait for draining to complete.
•
Remove both inlet flaps.
•
Remove a solution container.
Continued on next page
7-2
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...