4. Sample measurements
Calculation of FShunt and ctO
2
(a -
_
V
)
Calculation of
To obtain the calculated values of
F
Shunt and
c
tO
2
(a-v
–
), it is necessary to
analyze a mixed-venous and an arterial (or capillary) blood sample from the
patient.
FShunt and
ctO
2
(a-v
–
)
Do the following:
Step Action
Make, if required, a new patient report layout (see
Patient Reports
in
chapter 3
) with the following parameters included:
1.
•
F
Shunt and/or
c
tO
2
(a-v
–
) into the Patient Result
•
p
O
2
(v
–
),
s
O
2
(v
–
),
F
O
2
(I), RQ and
T
(patient temperature) into the
Patient ID
Analyze the mixed-venous sample and record the
p
O
2
(v
–
) and
s
O
2
(v
–
).
2.
3.
Analyze the arterial sample.
On the
Patient Identification
screen, select the sample type as
“Arterial“ or “Capillary“ and key in the following values:
4.
•
p
O
2
(v
–
) and
s
O
2
(v
–
) from the mixed-venous sample (step
2
above)
F
O
2
(I) if it differs from the default value of 0.21 (for
F
shunt)
•
RQ if it differs from the default value of 0.86 (for
F
shunt)
•
T
if it differs from the default value of 37 °C (for
F
shunt).
•
p
50(st) from the mixed-venous sample if
s
O
2
(a) > 97 %.
•
NOTICE:
The
F
Shunt value will be available as an estimated value even
without the data obtained from the mixed-venous sample.
4-25
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...