4. Sample measurements
ABL800 FLEX Operator's Manual
Introducing an expired air sample
Procedure
To analyze an expired air sample, do the following:
Step Action
1.
Open the syringe inlet flap and select the Expired Air mode.
NOTICE:
The Expired Air measuring program has to be selected in
the Analysis Setup for syringe samples in order to obtain the button on
the screen.
2.
Place the syringe tip in the inlet.
NOTICE:
The Expired Air mode is exclusively
used for samples of expired air. Injection of
blood or other liquid in this mode may cause
damage to the analyzer.
Press
Start
.
3.
4.
Inject the sample slowly, following the volume bar on the screen.
Make sure to continue injecting the expired air sample as long as the
volume is higher than 0 mL.
5.
When prompted by the analyzer status, remove the syringe and close the
inlet flap.
Enter the patient information on the
Patient Identification
screen.
6.
4-16