4. Sample measurements
ABL800 FLEX Operator's Manual
Introducing a blood sample
Syringe samples
To analyze a blood or other fluid sample without the use of the FLEXQ sampler
tray, do the following.
Step Action
1.
Check that the analyzer is in the Ready mode and scan the sample
barcode.
Accept or cancel the displayed patient data (selected in Sample Pre-
registration setup program – see
chapter 3
in this manual).
If no sample pre-registration was selected, proceed to step
2
below.
2.
Lift the syringe inlet flap. The following screen appears:
If sample pre-registration was selected (see
Analysis Setup
in
chapter
3
), press
Accept
to accept displayed patient data or
Cancel
to cancel
the patient data.
3.
CAUTION - Risk of incorrect result
Always remove air bubbles from a
sample before introducing it into the
analyzer. Air bubbles may result in
erroneous
p
O
2
values.
Place the syringe tip firmly in the inlet as
shown.
Safe
PICO sampler: Do not remove the
safe
TIPCAP.
4.
Select the desired mode by pressing the appropriate button.
Continued on next page
4-12
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...