4. Sample measurements
ABL800 FLEX Operator's Manual
Measurements with FLEXQ
Single
measurement
from the
FLEXQ sampler
tray
To make a single measurement, do the following:
Step Action
1.
Place the sampler in a slot on the tray. A short beep indicates that the
barcode has been accepted.
2.
The available measuring modes are shown; the selected mode is
highlighted. Change a measuring mode if desired.
Press
Proceed
.
3.
Enter, if required, additional patient data or mandatory data.
Press
Accept
to accept all the data (patient data, measuring mode,
selected parameters, layout, etc.).
4.
Remember to take the registration receipt printed out on the analyzer’s
printer.
Repeat steps
1-4
for another sampler.
5.
6.
The measurement starts. The patient name and the time left till the
result is ready are displayed on the
Ready
screen.
If further editing of the patient data has to be made, press the
button on the
Ready
screen to return to the
Sample Registration
screen.
7.
When the result is ready, it will be automatically transmitted to its
destination (the RADIANCE system or LIS/HIS).
Batch
measurement
To make a batch measurement, do the following:
Step Action
1.
Make sure that a slot indicator is green and ready to accept a sampler.
2.
Place the sampler in an empty FLEXQ sampler tray slot. A short beep
indicates that the barcode has been accepted.
3.
Make sure that your sampler is ready for processing. If not, check
analyzer connection to FLEXLINK.
4.
Proceed with the next sampler in the same manner as described above.
Continued on next page
4-10
Summary of Contents for ABL800 FLEX
Page 1: ...ABL800 FLEX operator s manual ...
Page 2: ......
Page 3: ...ABL800 FLEX operator s manual ...
Page 4: ......
Page 12: ......
Page 18: ...Contents ABL800 FLEX Operator s Manual ...
Page 196: ...5 Quality control ABL800 FLEX Operator s Manual 5 16 ...
Page 232: ...7 Replacements ABL800 FLEX Operator s Manual 7 26 ...
Page 282: ...10 Analyzer shutdown ABL800 FLEX Operator s Manual 10 10 ...
Page 366: ...11 Troubleshooting ABL800 FLEX Operator s Manual 11 84 ...
Page 428: ...15 Radiometer settings ABL800 FLEX Operator s Manual 15 22 ...
Page 436: ......
Page 438: ...Date of issue ABL800 FLEX Operator s manual ...