TitraLab 870 Reference Manual
Page 251
Turntable
blocked (SAC
error)
The turntable has been blocked or forced and or cannot function
properly.
Remove the obstruction and press
Resume analysis
(key
1
) to
continue the sequence from the point it stopped.
Warning!
Do not change the turntable before restarting the sequence from the
beaker it stopped (key
1
) or before restarting the sequence from the
next beaker (key
2
). The sample changer identifies a turntable only
when a new sequence is initialized (equivalent to a keystroke on
3
End of sequence
followed by
1
Run sequence
).
Type of method
The type of method is selected in the Add method menu while editing
a sequence. The following types are available:
•
Sample only methods,
•
Electrode calibration methods,
•
Reagent calibration methods,
•
Blank sample methods (Yes has been selected for Blank in the Edit
method menu),
•
QC sample methods (Yes has been selected for QC sample in the
Edit method menu).
User ID (Yes/No)
If specified, User ID will be asked for each time an analysis is run.
Enter in:
Setup menu > Configuration
User list
List of user defined methods, electrodes and reagents. The list is ini-
tialised to the preprogrammed methods, electrodes and reagents.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...