Titra
Lab
®
TitraLab 870 Reference Manual
Page 226
Sample unit
Unit of the sample amount. The sample unit can be expressed in
weight (g, mg) or volume (ml, µl) except for the ISE standard addition
method where volume (ml or µl) units are only available.
Enter in:
Edit method > Sample menu
Sample unit
conflict
In a Coupled method, the sample unit is not the same for the two titra-
tion methods.
Check the Sample screens of each method of the Coupled method.
Select electrode
Routine user
Is able to select an electrode to check the parameters and/or start a
calibration.
1.
Select the method/sequence using the electrode.
2.
Press
1
to start a calibration or
3
to check the electrode
parameters. In both cases, the electrodes available will be those
specified in the actual working method or sequence.
Supervisor
From the Electrode window.
1.
Press
4
.
2.
In the ID field, select the electrode from the User list.
Select method
To select a method (sample changer not in use):
1.
Select
Working mode = Method
in the Main window.
2.
Press
3
.
3.
In the ID field, select the method from the User list.
To select a method to be run using a Sample Changer:
1.
Select
Working mode = SAC Method
in the Main win-
dow. If this option is not available, enter the Setup menu >
Configuration and define a sample changer.
2.
Press
3
.
3.
In the ID field, select the method from the User list.
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...