TitraLab 870 Reference Manual
Page 211
Rinse burette
For safe reagent handling rinse the burette each time the burette is to
be replaced. It makes sure the delivery tubes and tubings are reagent
free. The rinse cycle lasts for approximately 5 minutes.
To rinse a burette:
1.
Enter Burette functions menu:
1.
Select the burette in the Address field,
.
2.
Press
4
Rinse.
The rinse procedure runs as follows:
•
Empties the burette.
•
Prompts you to remove the reagent bottle from the specified
address.
•
Fills the burette with air, then empties.
•
Prompts you to install the rinse bottle.
•
Runs the rinse programme using the rinse solution.
•
Prompts you to remove the rinse bottle.
•
Fills the burette with air, empties, then refills.
Rinse time
When a Sample Changer is in use enter the time (in minutes and
seconds) the electrodes should be immersed in each static rinse
beaker.
Refer to "Number of static rinses", page 166
Enter in:
Setup menu > Configuration (if
Sample changer
=
SAC80
,
SAC90
,
SAC850
or
SAC950
)
Range available:
00:00 to 30:59 min:s
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...