Titra
Lab
®
TitraLab 870 Reference Manual
Page 131
Flush burette
A flush rinses the burette, tubings and delivery tip thoroughly. It is
recommended to flush the burette on a daily basis.
The Titration Manager carries out a flush after each installation or rea-
gent replacement procedure.
To run a flush:
1.
Enter the Burette functions menu.
2.
Select the burette in the
Address
field.
3.
Press
3
Flush.
The flush procedure runs as follows:
•
Empty,
•
Fill 4 ml (*) then empty (5 times).
(*) 1 ml for a 10 ml burette and 10 ml for a 50 ml burette.
The length of the “Flush” procedure depends on the volume of the
burette.
You can also flush simultaneously all installed burettes,
"Global flush burettes", page 133
Format
(printouts)
Format = Listing
The whole report is printed in one operation.
Format = Page by page
The printer waits until a preset number of lines have been collected
then prints one page (this number is set by the
Nb line per
page
see "Nb lines per page (printouts)", page 161
The printing format applies for automatic printouts (at the end
of a test) or manual printouts (by pressing key
).
Access:
Setup Menu > Configuration
Summary of Contents for TitraLab 870
Page 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Page 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Page 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Page 23: ...Practical examples s...
Page 24: ......
Page 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Page 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Page 41: ...Glossary...
Page 42: ......
Page 255: ...Appendixes...
Page 256: ......
Page 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Page 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Page 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Page 286: ...TitraLab TitraLab 870 Reference Manual Page 286...