TEX2503TFT
25
/ 34
User Manual
Rev. 1.0 - 04/12/20
Menu 9
Stat
Enable (Remote) or disable (Local) the commands coming from remote.
Stat
Enables (x10) or disables (x1) the multiplication function of the
instantaneous modulation reading. This display mode is useful when
you want to view low levels of deviation.
PgD
Adjustment of the Power Good threshold relating to the forward power.
The percentage value of Power Good refers to the nominal power of
the device (2500 W for the
TEX2503TFT
), not to the forward power
delivered. So if you set a value equal to 50%, it will correspond to
1250 W, regardless of the power set. When the output power drops
below the set Power Good threshold value, the device changes the
state of pin [7] of the DB15 “Remote” connector on the rear panel.
PgR
Adjustment of the Power Good threshold relating to the reflected power.
The percentage value of Power Good refers to the nominal power of the
device (250 W for the
TEX2503TFT
respectively), not to the reflected
power delivered. So if you set a value equal to 5%, it will correspond
to 12,5 W respectively, regardless of the power set.
NOTE: This alarm does not move any contact on the DB15 “Remote” connector.
IIC
I
2
C network address setting, relevant when the exciter is connected
in an RVR transmission system that uses this protocol. However, it is
recommended not to modify it without reason.
IP
Display of the IP address assigned to the device (with
/ TLW-TEX-E
option).