background image

MODEL 61302L

BAROMETRIC PRESSURE SENSOR

INSTRUCTIONS

R.M. YOUNG COMPANY

2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN  49686, USA

TEL: (231) 946-3980    FAX: (231) 946-4772

PN: 61302L-90

 REV: H111215

1.0 SPECIFICATIONS

Pressure 

500 to 1100 hPa standard range

Digital Accuracy* 

0.2 hPa (25°C)

 

0.3 hPa (-40°C to +60°C)

Analog Accuracy**  0.05% of analog pressure range

Analog Temperature

    Dependence** 

0.0017%  of  analog  pressure 

range per °C (25°C reference)

Long Term Stability:  0.2% FS per year

  *  Defined as ±1 standard deviation from NIST-traceable 

pressure reference in clean, dry air.  Includes non-

linearity, hysteresis, repeatability, and calibration 

uncertainty.

 

  **  Defined as ±1 standard deviation from ideal analog 

output.  Total analog output accuracy is the root sum 

square of digital accuracy, analog accuracy, and analog 

temperature dependence.

Output Rate 

1.8 Hz (max) to 1 per minute

 

Current Output 

4 to 20 mA

 

1 2 - b i t   r e s o l u t i o n   ( 1   i n   4 0 0 0 )  

Selectable pressure range: 500 to 

1100 hPa (standard)

Serial Output 

Full-duplex RS-232

 

Half-duplex RS-485

 

1200 to 38400 baud

 

Continuous ASCII text

 

Polled ASCII text

 

NMEA

 

0.01 hPa resolution

Supply Voltage 

7 to 30 VDC

 

25 mA max in 4-20 mA mode              

 

7 mA with serial I/O only

 

Case 

Fiber-reinforced thermoplastic

Weight 

44 g (1.5 oz)

Reviews: