3.4 Environmental conditions

Place the instrument on a flat, horizontal surface with sufficient amount of space around 
it. Avoid setting up at places where it can fall down.

  Warning:
 

 

Please, pay attention that the pulled out drawer doesn’t reach out over the  

 

 

 

working place. This might lead to injuries of the employees and can damage  

  the 

instrument.

Avoid vibrations, strong electromagnetic waves (not computer), sources of heat and the 
exposure to direct sunlight.
Keep the workplace surrounding the instrument free of any aggressive liquids.

3.5 Environmental parameters

Operation temperature: 15 °C to 35 °C
Working height: 0 to 2.000 m above sea level
Relative humidity:  20 % - 80 %, no condensing

     Caution: 
     

Before turning the instrument on, acclimatize the instrument at room temperature  

     

to let all condensed water evaporate from the inside.

Summary of Contents for RIDA qLine Scan

Page 1: ...User Manual RIDAqLine Scan https clinical r biopharm com...

Page 2: ...2 RIDA qLine Scan User Manual...

Page 3: ...es no liability for errors or omissions or for any damages resulting from the application or use of this information The content of this manual is the property of R Biopharm AG and is copyrighted Exce...

Page 4: ...arameters 12 3 6 Transport and storage conditions 13 4 Installation and connection of RIDA qLine Scan 14 4 1 Transportation lock 14 4 2 Preparation 15 4 2 1 Connection 15 4 3 Installation 16 5 QC test...

Page 5: ...instrument or training requirements if in any doubt before using the instrument Consult User Manual where symbol left is seen on the instrument Warnings Cautions and notes Special notes in these oper...

Page 6: ...device Reference number Caution WEEE Waste Electrical and Electronic Equipment WEEE Waste Electrical and Electronic Equipment Refer to user manual Manufacturer Date of manufacturing RIDA qLine Scan s...

Page 7: ...e avoid any electrostatic charge Pay attention to all WARNING CAUTION and STOP statements in this publication Never open the electrical compartment while the equipment is plugged into a power source W...

Page 8: ...r professional use 2 2 Functions RIDA qLine Scan allows RIDA qLine Allergy tests to be scanned after manual processing The images are then quantitatively analyzed with RIDA qLine Soft Instrument featu...

Page 9: ...ications General weight 2 9 kg Plastic contents 2 parts of body housing Drawer mechanical drawer Power Supply 1 x external power supply Mod 3A 183WP24 Input 100V 240V 50 60 Hz 0 6 A Output 24V 0 75 A...

Page 10: ...tion This chapter describes how to install and start this device This includes unpacking accessories the software installation and PC connections Note Make sure you do not drop the scanner this could...

Page 11: ...positioning of the device Place the instrument on a flat horizontal surface Remove all indicated safeguarding and adhesive tape Put no heavy objects onto the reader Warning Please pay attention to th...

Page 12: ...es of the employees and can damage the instrument Avoid vibrations strong electromagnetic waves not computer sources of heat and the exposure to direct sunlight Keep the workplace surrounding the inst...

Page 13: ...to 95 Storage Environmental temperature 20 C to 30 C Relative humidity 20 to 30 It is recommended to choose the transport and storage conditions in the medium range of the conditions specified above...

Page 14: ...livered in a locked state The dark screw locks the scanning unit whereas the light colored screw seals a second opening In order to unlock the device please unscrew both screws completely remove the n...

Page 15: ...ocket at the rear of the scanner Use only the external power adapter provided with the instrument Caution Only use the original power adapter of the RIDA qLine Scan The plugs cannot be mixed up due to...

Page 16: ...The plugs fit into the dedicated sockets only and you cannot mix them up Please switch on the RIDA qLine Scan You will find the switch at the rear of the device The small blue lamp on the right front...

Page 17: ...nnected PC To perform the QC test click the down arrow in Settings menu item and select Perform QC test Place 10 QC strips Art No ZG1108 in the strip holder included in the RIDA qLine Incubation Set I...

Page 18: ...ow the results for each individual strip 6 Image creation with RIDA qLine Scan 6 1 Placing the test Pull out the drawer at the front side of the RIDA qLine Scan Open the drawer until you see the compl...

Page 19: ...ning please check if the membranes were evaluated with good quality Repeat the scan if necessary 8 Turn off the reader At the rear of the device you can find a switch to turn the device on and off Whe...

Page 20: ...de The cleaning of the interior is limited to the drawer area Warning Wear rubber gloves when cleaning the scanner Slight contaminations Pull out the drawer and then wipe the drawer with a damp cloth...

Page 21: ...s procedure until the glass sheet is clean see picture Put the device back to its initial position carefully Please place the drawer back into its guideways and close it completely Finally reconnect t...

Page 22: ...ers and end users in countries of the European Union EU must not dispose of electronic and electrical equipment as well as electronic and electrical accessories in household waste Within the EU please...

Page 23: ...hat the RIDA qLine Scan is compliant with Regulation EU 2017 746 In Vitro Diagnostic Medical Devices In addition the RIDA qLine Scan fulfils the EMC protection requirements according to EN 61326 2 6 T...

Page 24: ...24 RIDA qLine Scan User Manual R Biopharm An der neuen Bergstra e 17 64297 Darmstadt Germany E mail orders r biopharm de clinical r biopharm com Version 2022 04 0...

Reviews: