17
4.7.1 Lubricate (grease) the Probe
O-Ring
Apply a small amount (approx 1mm ‘blob’) of
silicone grease to the o-ring - illustrated in Figure 1.
Place a 1000 mL tip onto the tip-picking probe, rotate
the tip on the probe twice over in order to spread the
grease over the whole o-ring, then remove the tip.
Thoroughly wipe the metal parts of the probes
(and not the o-rings) clean with tissue, so that there
is no grease left on the probes themselves.
4.7.2 Mount and Connect the Touchscreen
Figure 1:
probe o-ring
Figure 2:
touchscreen rear clamp
mounting
Figure 3:
touchscreen-to-instrument
mounting