5 Information required by the user
5.1 Explanation of the signal lights
Tab.1
Explanation of the signal lights
Activity of the signal lights
Meaning
All signal lights flash simultaneously
Error, software shows error message
No signal light lights up
•
Device is off
•
Device is waiting for input
•
Device is defective
All signal lights light up briefly one after the
other
Self-test after switching on the device
Signal light of the respective measuring
channel flashes/illuminates continuously
Calibration/measurement in progress
5.2 Notes on troubleshooting
If there is an error, all four signal lights flash simultaneously. In this case, check the software
to see which error is present.
In the event of the following errors, a new measurement can be carried out immediately after
the error has been corrected:
Tab. 2
Type of error
Causes and correction of errors
Calibration failed
During calibration, there was a microtiter plate in the
slot: Check if there is a microtiter plate in the slot and
remove it before the instrument performs calibration.
The inside of the slot is dirty: Clean the inside of the slot
according to the cleaning instructions in chapter 6.2.
Summary of Contents for RIDA ABSORBANCE 96
Page 1: ...R Biopharm AG...
Page 2: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 2...
Page 4: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 4 Bedienungsanleitung...
Page 31: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 31...
Page 32: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 32 User manual...