Swapping Sensors on Systems Equipped
with the Video Camera Option
This subsection only applies to Cobra systems equipped with the video camera
option, and describes how to remove the video camera assembly and sensor so
that you can swap sensors.
Note:
The video camera option can only be used with DRS-300 and DRS-500
lasers; it cannot be used with the DRS-2000 or DRS-8000 lasers.
Removing the Video Camera Assembly and Sensor
To remove the video camera assembly and sensor, follow the steps below. You
will need 2 mm and 3 mm hex key (Allen) wrenches.
1.
Power down your Cobra.
Never connect or disconnect the laser sensor cable from the sensor while power
is applied to Cobra. Doing so could seriously damage the sensor.
2.
Disconnect the laser sensor cable from the top of the sensor.
3.
Use a 2 mm hex key (Allen) wrench to remove the four screws securing the
camera cover to the video camera assembly (see Figure 2-4 on page 2-6).
4-2
Swapping Sensors
Summary of Contents for Cobra DRS-300