QM1004A-2-18 Combined RF Upconverter/Downconverter
User Manual
6.6 SYSTem:READstate? [0-5]
This query allows the user to read the parameters of the stored states without changing any internal reg-
isters. The query output is a list of parameters for the selected state. If no state parameter is given, the
returned state is state 0. The list of state parameters contains comma separated values, without spaces,
in the following order:
Tune Frequency, RF Power State, LNA Power State, Reference Frequency, Exter-
nal Reference Enabled, External Referece Switch Override, PLL Mode, Reference Divider, Upconverter
Attenuation, and Downconverter Attenuation
. See the respective command definitions for descriptions of
each parameter.
Syntax
:
SYST
:
READ
?
Space
0-5
Query
:SYST:READ? 0
This query requests the parameter values of state 0.
Response
: 10.0000,0,0,10,0,0,1,1,0,0
∙
10.0000 = FREQuency:TUNE is set to 10 GHz. (page 45)
∙
0 = POWEr:RF is set to 0 or OFF. (page 50)
∙
0 = POWEr:LNA is set to 0 or OFF. (page 49)
∙
10 = FREQuency:REFerence:FREQuency is set to 10 MHz. (page 43)
∙
0 = FREQuency:REFerence:EXTernal is set to 0 or Internal. (page 42)
∙
0 = External Reference Switch Override is 0 (switch-controlled).
∙
1 = FREQuency:PLLMode is set to 1 or Integer. (page 39)
∙
1 = FREQuency:REFerence:DIVider is set to 1. (page 41)
∙
0 = POWEr:CH1ATTEN is set to 0 dB (page 47)
∙
0 = POWEr:CH2ATTEN is set to 0 dB (page 48)
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is less than 0 or greater than 5, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 1.4.3
66