22
Installing a Stand-Alone Control Module
Testing the Digital Level
Required tools: digital level
Perform this test before each use of the digital level and any time the digital level has been dropped or is
being used in an environment that varies ±9
°
F from the environment in which it was last calibrated. If the
digital level fails the accuracy test you must re-calibrate the level before use.
1
Turn on the level by pushing the
ON/OFF
button. Position the level with the display facing you and the
text on the face of the level right-side up. Ensure that the level is on a clean, flat, and horizontal surface.
This surface does not have to be exactly level. Wait 10 seconds for the level to completely settle and
take note of the angle on the display.
2
Rotate the level end-for-end so the display is facing away from you. The screw on the back of the level
should be on the left side. Be sure to set the level in the exact spot as in Step 1. Wait 10 seconds for
the level to completely settle and take note of the angle on the display.
3
Roll the level toward you so that the display is facing you, but the lettering on the face of the unit is
upside down. Be sure to set the level in the exact spot as in Step 1. Wait 10 seconds for the level to
completely settle and take note of the angle on the display.
Summary of Contents for Scalar i6000
Page 20: ...8 About This Guide and Your Product...
Page 38: ...26 Installing a Stand Alone Control Module...
Page 104: ...92 Installing a Multi Module Library...
Page 156: ...144 Installing Cartridges...
Page 164: ...152 Setting up Your Library for Access...
Page 242: ...230 Configuring the Library...
Page 304: ...292 Adding Optional Hardware eight 2 5 mm screws...
Page 318: ...306 Adding Optional Hardware...
Page 336: ...324 Installation Testing and Verification Figure 35 Example Test Log Output...
Page 356: ...344 Testing and Calibrating the Digital Level...
Page 362: ...350 LBX Board and Terminator...
Page 380: ...368 Glossary...
Page 384: ...372 Index...