Chapter 1 Specifications
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
1-19
SYSTEM OPTIONS
The following lists the available options for HF Series X-Ray Generators. Note that
some options, such as QG-150 and Q-HSS, may only be used on certain generator
models (refer to Tables 1-5 through 1-7 for compatibility):
•
R80-HS: Exposure Hand Switch and R80-FS Foot Switch
•
QG-PDL: Operator Control Panel Pedestal Mount (used with ODYSSEY
control panel)
•
QG-WM: Operator Control Panel Wall Mount (used with ODYSSEY control
panel)
•
QGV-WM: Operator Control Panel Wall Mount (used with Q-VISION control
panel)
•
QG-150: 150 kV High Voltage Transformer (standard equipment on Models
QG-65 and QG-80)
•
Q-HSS: Internal High (Dual) Speed Starter Unit (standard equipment on
Models QG-65 and QG-80
•
Q-Connect: Enables ethernet connection to a PC computer using Quantum’s
"QBus" network integration software; required on systems utilizing a touch-
screen operator control panel
•
QG-28VPS: Internally mounted Aux28 VDC/8A Power Supply (only
available with three-phase input generators)
•
QG-TOMO: Provides interface for operation with Siemens and Pausch
tomographic systems (not available with Q-VISION HF Series OCP’s)
•
TechVision: Supplementary x-ray control panel, mounts on tube support
device. Refer to the TechVision Remote Control Panel Service and
Operator’s Manuals for complete installation and operation instructions.
NOTE: The use of ACCESSORY equipment not complying with the
equivalent safety requirements of this equipment may lead to
a reduced level of safety of the resulting system. Consider-
ation relating to the choice shall include:
-use of the accessory in the PATIENT VICINITY
-evidence that the safety certification of the ACCESSORY has
been performed in accordance with the appropriate IEC 601-1
and/or IEC 601-1-1 harmonized national standard.
SYSTEM OPERATING ENVIRONMENT
Ambient Temperature:
+10°C to +40°C
Relative Humidity:
30 to 75%, non-condensing
Atmospheric Pressure:
700 hPa to 1060 hPa
Typical Heat Output of
Generator Cabinet:
3600 BTU per hour
Typical Heat Output of
operator control panel (OCP): 130 BTU per hour
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...