Chapter 1 Specifications
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
1-7
NOTE: All single-phase x-ray generator input ratings assume the use of a
single-phase distribution transformer. When utilizing two legs of a 3-
phase transformer, review the National Electrical Code and any appli-
cable local electrical code(s) to ensure the correct rating distribution
transformer (in terms of KVA) is used to assure line stability. This
configuration would require an increase in the overall power (KVA) of
the distribution transformer, depending on whether the input line
transformer is Wye or Delta type. Under no circumstances should a
Delta type transformer be used for input power to a 3-phase genera-
tor.
SINGLE-PHASE X-RAY GENERATORS
Non-Stored Energy Single-phase x-ray generators require three-wire line inputs in
either of the following two configurations:
•
one hot wire, one neutral wire, and one earth ground wire
•
two hot wires and one earth ground.
Refer to Table 1-1 for power requirements of single-phase (non-stored energy)
HF Series Generators.
25 kW Single Phase (Model QG-25)
Line Voltage:
208-260 VAC (±5%), 100 Amps, 50/60 Hz
Momentary Current:
200 Amps (@208VAC)
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
25 kW Single Phase (Model QG-25-5)*
Line Voltage:
380 - 480 VAC (±10%), 60 Amps, 50/60 Hz
Momentary Current:
110 Amps @ 415 VAC
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
32 kW Single Phase (Model QG-32)
Line Voltage:
208-260 VAC (±5%), 120 Amps, 50/60 Hz
Momentary Current:
240 Amps (@208VAC)
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
32 kW Single Phase (Model QG-32-5)*
Line Voltage:
380 - 480 VAC (±10%), 75 Amps, 50/60 Hz
Momentary Current:
140 Amps @ 415 VAC
Long Term Current:
less than 1.0 Amp
Mode of Operation:
Intermittent operation
Summary of Contents for ODYSSEY HF Series
Page 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Page 15: ...Chapter 1 1 SPECIFICATIONS 1...
Page 16: ...1 2...
Page 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Page 38: ...2 2...
Page 93: ...Chapter 3 1 CALIBRATION 3...
Page 94: ...3 2...
Page 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Page 166: ...4 2...
Page 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Page 186: ...5 2...
Page 281: ...Chapter 6 1 DIAGRAMS 6...
Page 282: ...6 2...
Page 349: ...A 2...
Page 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Page 365: ...B 2...
Page 420: ...C 1 Log4View Error Logger C Appendix...
Page 421: ...C 2...
Page 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Page 429: ...D 2...