36
CANopen
HBM
QuantumX
Input scaling 1, field value (electrical input, float format, point 1)
Object description
Index
x120 (hex)
24864 (dec)
Note
Object name
AI_Input_scaling_1_FV
This object defines the field value of the first
calibration point for the analog input channel
Object code
Array
calibration point for the analog input channel.
It is scaled in the physical unit of the field
Data type
Float, INT16, INT24, INT32
It is scaled in the physical unit of the field
value.
CiA standard
Yes
Category
Optional
Value description
Sub
-
index
0 (hex)
1 (hex)
2 (hex)
to
C7 (hex)
Description
Number of entries
Input scaling 1 FV
1
Input scaling 1 FV 2
Input scaling 1 FV
199
Entry category
Mandatory
Mandatory
Optional
Optional
Access
ro
rw
rw
rw
PDO mapping
No
No
No
No
Value range
1
−
254
Data type specific
Data type specific
Data type specific
Default value
No
No
No
No
Input scaling 1, process value (physical output, float format, point 1)
Object description
Index
x121 (hex)
24865 (dec)
Note
Object name
AI_Input_scaling_1_PV
This object defines the process value of the
first calibration point for the analog input
Object code
Array
first calibration point for the analog input
channel. It is scaled in the physical unit of
Data type
Float, INT16, INT24, INT32
channel. It is scaled in the physical unit of
the process value.
CiA standard
Yes
Category
Optional
Value description
Sub
-
index
0 (hex)
1 (hex)
2 (hex)
to
C7 (hex)
Description
Number of entries
Input scaling 1 PV 1
Input scaling 1 PV 2
Input scaling 1 PV 199
Entry category
Mandatory
Mandatory
Optional
Optional
Access
ro
rw
rw
rw
PDO mapping
No
No
No
No
Value range
1
−
254
Data type specific
Data type specific
Data type specific
Default value
No
No
No
No
Summary of Contents for CX27