MINISYSTEM Manual Version 1.5
Page 23
a. High Power
– The laser energy will range from 0 to the value set with the calibration
command. (See ECS in the command section).
b. Low Power
– The laser energy will range from 0 to 25% of the value set with the
calibration command. (See ECS in the command section).
Cut Size
The cut size value is measured at the sample. Example: a 2500µm aperture opening with a
50x microscope objective has a cut size of 50.00µm.
1. Magnification
– specified magnification of microscope objective being used.
2. X-size & Y-size
– value corresponds to image size which is determined by the aperture
size (range 0 to max size) divided by the microscope objective magnification. The max
size of the aperture is specified at time of order. The system ID? query will tell you what
size slit is installed on the system.
3. Angle
– Values are ± range of 0 to 450 (45.0 degrees). Positive values rotate the
aperture positive, Negative values rotate the aperture negative
. A “-“ must be entered for
the aperture to rotate in the negative direction.
Objective
Magnification
Max Size for
2.5mm Slit
(mm)
Max Size for
4mm Slit
(mm)
1x
2500
4000
2x
1250
2000
5x
500
800
10x
250
400
20x
125
200
50x
50
80
100x
25
40
200x
12.5
20
Illumination
The MINISYSTEM is configured with an illuminator used to show the XY-Theta aperture.
1. Backlight Level
– values are 0 to 1000 (0 to 100.0%).