Coherent Modulation Transmitter | IQTX
Quantifi Photonics Ltd.
Version
1.04
12
Separating a connection
When separating a connection:
•
Support the devices to avoid any twisting, rocking, or bending force on either connector.
•
Use an open-end wrench to prevent the device body from turning.
•
Use another open-end wrench to loosen the connector nut.
•
Complete the disconnection by hand, turning only the connector nut.
•
Pull the connectors straight apart.
Using a torque wrench
Applying the correct torqueing force by using a torque wrench avoids damage due to over or under
tightening connections. It significantly improves measurement accuracy and repeatability at mm-wave
frequencies, avoids signal failure and extends connector life.
When using a torque wrench:
•
Make sure the torque wrench is set to the correct torque setting (refer 4.5 Torque settings).
•
Position torque wrench, and a second wrench to hold the device or cable, within 90° of each
other before applying force. Make sure to support the devices to avoid putting stress on the
connectors.
•
Hold the torque wrench lightly at the end of the handle then apply force perpendicular to the
torque wrench handle. Tighten until the “break” point of the torque wrench is reached. Do not
push beyond initial break point.