Automated Optical Switch | Switch 1000/1100/1400 Series
Quantifi Photonics Ltd.
Version
3.04
13
5
Handling the Switch 1000/1100/1400 Series products
Switch 1000/1100/1400 PXIe module installation
STEP 1:
Power OFF the Chassis
STEP 2:
Remove the module
from the antistatic bag.
Retain bag
STEP 3:
Align module with slot
guide rails
STEP 4:
Push module into slot
until resistance is felt from the
backplane connection
STEP 5:
Engage the
fastening clip. Secure all
fastening screws
STEP 6:
Power ON the chassis
•
Do not remove the Switch 1000/1100/1400 Series product from the antistatic packaging
until instructed during the following installation procedure.
•
The Switch product is sensitive to ESD. Ensure you are wearing a grounded wrist strap
at all times when handling the Switch product to prevent such damage.
•
Take care not to handle the connectors on the Switch product, as once they are exposed
to skin contact this may leave corrosive residue which can damage the connector.
DO NOT attempt to remove or adjust any component of the PXIe chassis while the power is
on. Ensure the chassis is powered OFF, and that the correct handling procedure detailed
herein is followed when removing or installing any modules.
After powering on the PXIe chassis, please wait at least 2 minutes before attempting to
communicate with the instrument. This will allow the chassis enough time to finish boot
procedures and initialize the communication server.