Environmental Conditions:
•
Ambient temperature range: 10°C to 40°C for specified system performance
specifications.
•
Storage Temperature Range: 5°C to 50°C.
•
Altitude: 3048m (10,000 ft) maximum altitude.
Regulatory Compliance:
•
Complies with the Restriction of Hazardous Substances Directive (RoHS)
2002/95/EC.
•
Electromagnetic Compatibility (EMC).
•
Conforms to EMC Directive 89/336/EEC and Low Voltage Directive 73/23/EEC.
•
Refer to the
standards tested relating to electromagnetic emissions and immunity.
Safety:
•
Conforms to EN 61010-1 and IEC 61010-1 general safety requirements for
electrical equipment for measurement, control, and laboratory use.
•
Installation/over voltage category II; Pollution degree; 2 Standard
environmental rating.
•
Enclosure degree of protection rated IEC 60529 IPX0.
•
Conforms to Laser Emission Equipment standard EN 60825-1 and Federal
Laser Product Performance Standard, 21 CFR Part 1040.10 (FDA-CDRH)
except for deviations pursuant to CDRH Laser Notice 50.
Installation:
Standard ICE450 operates in upright or horizontal position
•
Dimensions (H x W x L): 14” x 5¼” x 18” [35.6cm x 13.3cm x 45.7cm]
•
Weight: 29.0lbs [13.2kg] without coolant; approx. 31.6lbs [14.3kg] with
coolant.
19-inch 3U Rack ICE450 used for equipment racks
•
Dimensions (H x W x L): 5¼” x 19” x 20¼” [13.3cm x 48.3cm x 51.4cm].
•
Weight: 28.6lbs [13.0kg] without coolant; approx. 30.2lbs [13.7kg] with
coolant.
TECHNICAL SPECIFICATIONS
Quantel
reserves the right
to modify the
specifications
without notice.
1.0 General Specifications
Power input:
•
100-240VAC, 50/60Hz, 850VA; Power Factor Corrected (PFC)
•
AC mains supply voltage shall not fluctuate more than ±10% of nominal voltage.
•
User serviceable power input fuses, both lines fused: 10A, 250V, Time Delay,
5mm x 20mm, and Ceramic Tube.
•
Universal type IEC 320 power line connector for detachable cord.
Cooling System:
•
Air-cooling of the Power Supply is required. Airflow is front to back.
•
Coolant: Distilled water or Reagent Grade Ethylene Glycol/distilled water mixture.
•
Coolant temperature regulation—the system is designed to regulate to 38°C,
with input power sufficient for the ambient conditions present, regardless of
the input power. Coolant temperature regulation depends on factors such as
ambient temperature, operating voltage, and the average operating power of
the laser (which may vary from about 65W up to 200W).
Output Power Operating Voltage Ambient
Temperature
Measured
Temperature Rise
150W
230VAC
28°C
10.2°C
200W
230VAC
28°C
10.7°C
300W
230VAC
28°C
13.5°C
•
De-Ionizing (DI) cartridge inline with the ICE450 coolant lines maintains
coolant conductivity at 1.0μS m-1 (resistivity 1.0M
Ω
m).
•
DI cartridge contains a particle filter.
•
Coolant connectors are shut-off quick-disconnect style.
•
Coolant vent and fill port are accessible from the front panel for filling and
draining the coolant.
•