IV - Maintenance

 

 

 

 

XL_FUSION_ME_AN_160314

 

 

1

 

PREVENTIVE MAINTENANCE 

Contact your QUANTEL MEDICAL or your local distributor at least every six months to carry out a preventive 
maintenance visit to confirm the correct function of your system. 
 
At each six-monthly visit QUANTEL MEDICAL or your local distributor will clean the optics, check the safety 
filter coatings, and check general performance, alignment and calibration of the system. 
 
At least once a year QUANTEL MEDICAL or your local distributor will fully re-calibrate the system and check 
ground resistance and earth leakage current (in accordance with IEC 60601-1). 

 

2

 

CARE AND MAINTENANCE OF THE OPTIMIS FUSION LASER SYSTEM 

 

2.1

 

CLEANING AND DISINFECTION FOR PATIENT AREAS 

The use of the OPTIMIS FUSION involves only brief skin contact with the patient, but attention should be 
given to the possibility of cross-contamination between patients via the patient contact areas of the system. 
These are the chin rest, headrest and steadying handles. 
 

2.1.1

 

RESPONSIBILITIES FOR CLEANING AND DISINFECTION 

The health care facility where the instrument is to be used is responsible for: 

1)

 

Determining the level of cleaning and disinfection of patient contact areas required between patients. 

2)

 

Appropriate education and training for staff required to carry out cleaning and disinfection. 

3)

 

Ensuring  that  routine  cleaning  and  disinfection  methods  used  in  the  facility  are  compatible  with  the 
instrument. 

4)

 

Routine cleaning of the entire system. 
 

2.1.2

 

CLEANING AND DISIFECTION OF PATIENT AREAS 

Although cleaning and disinfection procedures and standards within different health care facilities may vary 
considerably, the following points are provided as general guidelines: 

1)

 

Thorough cleaning of all patient contact areas is recommended for each procedure.  

2)

 

Disposable chinrest papers can also be attached to the chinrest area and changed between patients.

 

3)

 

Manual cleaning can be carried out by wiping all contact areas using a suitable liquid cleaning agent 
which is non-corrosive, non-toxic and Iow in residue. 

 

2.1.3

 

CLEANING THE GONIOSCOPY CONTACT LENSES 

 

The contact lenses for use with the OPTIMIS FUSION have a special low-reflectivity coating bonded to 
the lens, and they must therefore be handled carefully. 
As soon as a lens is removed from the patient’s eye, thoroughly rinse it in cold or warm water to remove 
salts, mucous and gonioprism solution. Wash in warm water with a few drops of clear dishwashing liquid, 
then rinse with cool water and blot dry. Dry completely before storing in the case. 
To effectively disinfect the lens, soak in a solution of sodium hypochlorite 1,2° chl. for a recommended 
exposure time of 30 minutes. Rinse thoroughly with cool water and blot dry. Dry completely before storing 
in the case. 
 
To  sterilize,  use  ethylene  oxide  gas  with  aeration  not  to  exceed  125°F  (52°C),  following  the  sterilizer 
manufacturer’s recommendation. Remove the lens from the case before sterilizing. 

 

 

CAUTION: 
Do not boil or autoclave a lens. Do not use alcohol, hydrogen peroxide or acetone on a lens, as this will 
damage it and render it unusable. 
 
 
 
 

 