QuadTech Guardian 2500 Series Instruction Manual Download Page 1

 

 
 

 

                               

Guardian 2500 Series 

AC/DC/IR Hipot Tester 

                                 Instruction Manual 

 

 

      

 

 

 

 

   Form 150319/A8 

 
 

  

QuadTech, Inc., 1999 

 

Domestic Sales             800-253-1230 

 

For additional contact information please visit our website  

www.quadtech.com

 

 
 
 

The material in this manual is for informational purposes only and is subject to change, without notice.  
QuadTech assumes no responsibility for any error or for consequential damages that may result from the 
misinterpretation of any procedures in this publication. 

 
 

WARNING 

Potentially dangerous voltages may be present on front and rear panel terminals.  Follow all warnings in 

this manual when operating or servicing this instrument.  Dangerous levels of energy may be stored in 

capacitive devices tested by this unit. 

Always make sure the high voltage indicator is 

not

 on when connecting or disconnecting the device under 
test.

 

 
 

 

Product will be marked with this symbol (ISO#3684) when it is necessary for the user to refer to 

the instruction manual in order to prevent injury or equipment damage.

 

 

   

Product marked with this symbol (IEC417) indicates presence of direct current. 

 

  

 

Product  will  be  marked  with  this  symbol  (ISO#3684)  when  voltages  in  excess  of  1000V  are 

present.

  

 

 

 

 

Summary of Contents for Guardian 2500 Series

Page 1: ...lly dangerous voltages may be present on front and rear panel terminals Follow all warnings in this manual when operating or servicing this instrument Dangerous levels of energy may be stored in capac...

Page 2: ...Page 2 of 2...

Page 3: ...s Introduction Section 1 1 1 Unpacking Inspection 29 1 2 Product Overview 29 1 3 Front Panel Description 30 1 4 Rear Panel Description 34 1 5 Installation 35 1 5 1 Instrument Positioning 35 1 5 2 Clea...

Page 4: ...2 4 6 Beep On Key Hit 55 2 4 7 RS 232 Interface Optional 55 2 4 8 IEEE 488 Interface Optional 55 2 4 9 Enhanced Features 56 1 Test Low Limit 56 2 Delay Time 57 3 Continue Voltage On Fail 58 4 Continue...

Page 5: ...3 4 Enhanced Remote I O Interface 76 3 5 IEEE 488 2 Interface 78 3 5 1 General 78 3 5 2 IEEE 488 Rs 232 Commands 79 3 5 3 Remote Calibration 87 3 5 4 Formats 88 3 5 5 IEEE 488 Sample Program Quick Bas...

Page 6: ...Page 6 of 6...

Page 7: ...hanges in the Product not approved by QuadTech shall void this warranty QuadTech shall not be liable for any indirect special or consequential damages even if notice has been given of the possibility...

Page 8: ...Page 8 of 8...

Page 9: ...steps Voltage Display Accuracy 1 of reading 5V Resolution 1V steps DC Current Display Range 0 1 A to 1 0mA DC in 0 1 A steps 1 0mA to 7 5mA DC in 1 A steps Accuracy 0 5 of set high limit 0 1 A or 1 A...

Page 10: ...1 to 999sec 50ms DC Hipot Ramp Dwell Test Fall 0 1 to 999sec 50ms IR Ramp Dwell Test Fall 0 1 to 999sec 50ms GC Test 0 1 to 999sec 55ms Test can be set to Continuous Ramp Dwell Fall can be set to Off...

Page 11: ...Dimensions w x h x d 17x5 25x16in 432x133x406mm Weight 29 lbs 13kg net 36 lbs 16kg shipping Environmental Meets MIL T 28800E Type 3 Class 5 Operating 0 to 40o C Storage 10 to 60o C Humidity Maximum RH...

Page 12: ...Page 12 of 12...

Page 13: ...Continuity Cable labeled w QuadTech part 630114 Ground Return Cable labeled w QuadTech part 630115 High Voltage Lead Set 2 Meters consisting of 3 cables 2000 04 High Voltage Cable labeled w QuadTech...

Page 14: ...right are used as dress panels to adapt the 2500 to the standard 19 inch rack width THESE FLANGES SHOULD NOT BE USED AS SOLE MOUNTING SUPPORT OF THE 2500 in rack mount applications Chassis guides or o...

Page 15: ...ent capability of devices with high resistance and substantial parallel capacitance devices such as wire wound heaters ac inlet filters sensors etc When this option is installed additional ramp and dw...

Page 16: ...Page 16 of 16...

Page 17: ...voltage being applied to the output terminals 6 When the HIGH VOLTAGE light is flashing or constantly lit on some later units NEVER touch the device under test the lead wires or the output terminals...

Page 18: ...Page 18 of 18...

Page 19: ...hree wire grounded outlets 5 Apply power to the Guardian 2500 by pressing the front panel POWER switch to ON 1 position 6 The instrument performs a self test and will momentarily display the instrumen...

Page 20: ...time of 7 seconds Note the message Idle in the upper right display this area always indicates the instrument status Pressing the up down arrow key will also sequence through the test setups up or dow...

Page 21: ...nditions 1 Press the TEST NUMBER key to sequence up through the test setups selecting one for change or one not presently programmed as noted by the message No Test Programmed Continue to press the ke...

Page 22: ...mming parameters and arrows to make changes Display returns to Idle mode after programming the last parameter or Press arrow to reverse the selection of parameters or Exit Program mode at any time by...

Page 23: ...e and positioned in the approximate location of the expected test See Note below Test Lead Configuration for Measuring Offset 2 Press the UTILITY key to select the measure offset function Press up dow...

Page 24: ...s Part 2000 02 Connection Using Corded Product Adapter Part 2000 13 Black Lead Small White Lead G C RETURN HV OUT DUT Guardian 2510 20 30 Case Large White Lead White Banana to clip Black lug Black ban...

Page 25: ...nd press START green button to initiate the test The HIGH VOLTAGE indicator flashing or constantly lit on some later units serves as warning to the operator that high voltage is present The STOP butto...

Page 26: ...asurement exceeded high limit during ramp time Lo Fail Measurement was below low limit during test time Hi Fail Measurement exceeded high limit during test time Arc Fail Measurement exceeded arc limit...

Page 27: ...be initiated reset before another test can be initiated NOTE If Continue Voltage On Fail is enabled a Fail on the summary display as shown for Test 1 above can be a conditional Pass A conditional pas...

Page 28: ......

Page 29: ...real or imaginary current can be monitored and displayed The Guardian 2520 AC DC Hipot Tester has all the features of the Guardian 2510 with the addition of DC Hipot The test voltage can be programme...

Page 30: ...iption Figure 1 1 shows the controls and indicators on the front panel of the Guardian Series 2500 Table 1 1 identifies them with descriptions and functions Figure 1 1 Front Panel Controls and Indicat...

Page 31: ...est results 7 TEST Pressing this key will sequentially scroll through test setups in ascending order based on how the Utilities Menu is configured shows all tests or only those enabled Test number is...

Page 32: ...Off On Step And Increment Off On Clear All Tests No Yes Test To View All Tests Only Sequence Only Single Step Only Factory Perform Calibration Cal Date Start Full Start GC Only Ground Fault Stop Off...

Page 33: ...ts serves as VOLTAGE a warning to the operator that high voltage is present to the High Voltage Indicator Output connector 15 GROUND Continuity test connection measured between this connector and the...

Page 34: ...Page 34 of 34 Introduction 1 4 Rear Panel Description Figure 1 2 Rear Panel View AUXILIARY G C RETURN HV OUT IEEE 488 REMOTE I O RS232 ENHANCED REMOTE I O 1 2 RS232 3 5 4 6 REMOTE I O RS232...

Page 35: ...el connectors HIGH VOLTAGE Connections GROUND RETURN and GROUND CONTINUITY 3 REMOTE Standard Interface on all units except when ENHNACED REMOTE I O RS232 option is ordered Refer to paragraph 3 2 4 IEE...

Page 36: ...instrument can be damaged if the wrong fuse is installed To change the fuse refer to paragraph 1 5 4 1 5 4 Safety Inspection Before operating the instrument inspect the power inlet module or the rear...

Page 37: ...e fuse drawer This tray contains the active fuse left side secured by holder and spare fuse on the right side if present Refer to Figure 1 4 Remove the active fuse from the holder by prying upward usi...

Page 38: ......

Page 39: ...ent is to be used only with three wire grounded outlets Power is applied to the Guardian 2500 by pressing the front panel POWER switch to ON 1 position WARNING DO NOT TURN INSTRUMENT POWER ON OR OFF W...

Page 40: ...t 0 001mA to 15mA 0 0001mA to 7 5mA 0 01M to 2 000T Off 0 01 to 10 00 5 Lo Limit 0 001mA to 14 999mA Off 0 0001mA to 7 4999mA Off 0 01M to 1 999T 0 01 to 9 99 Off 6 Arc Limit 0 5mA to 15mA Off 0 5mA t...

Page 41: ...nuity Test Leave the display set for the selection desired and press right arrow key to move to the next parameter or PROGRAM key to exit program mode 2 To set Voltage Press up down arrow keys to incr...

Page 42: ...for previous parameter or PROGRAM key to exit program mode 4 To set Lo Limit Press up down arrow keys to increase or decrease the limit setting between Off and 0 001mA to 14 999mA for AC Current Off...

Page 43: ...or decrease the ramp time between Off and 0 1 to 999 9sec The test voltage increases linearly during the programmed ramp time The standard Hi Lo Limits are not checked during this ramp time for curren...

Page 44: ...the dwell time between Off and 0 1 to 999 9sec for DC Current and Insulation Resistance No Hi Lo Limits are checked during the programmed dwell time Press right arrow key to move to the next paramete...

Page 45: ...s to toggle the frequency between 50 or 60Hz Press right arrow key to move to the next parameter left arrow for previous parameter or PROGRAM key to exit program mode 13 To set Gnd Continuity Press up...

Page 46: ...Sequence Testing S1 S9 Test Programmable Tests for the Sequence 1 none 1 thru 25 skip 2 none 1 thru 25 skip calc 3 none 1 thru 25 skip calc Programmable Parameters for calc Programmable Hi lo Limits f...

Page 47: ...ce Press up down arrow keys to select the second test in the sequence as none skip 1 25 or calc see discussion of the calculate function step 4 below S1 2 Program Tests none S1 1 Program Tests none S1...

Page 48: ...or calc see discussion of the calculate function step 4 below Press right arrow key to exit the program mode and display the complete sequence for sequence test number 1 Program message now indicates...

Page 49: ...ed High Limit and then the up down arrow key to increase or decrease the limit between 0 001mA and 15 000mA Press the right arrow key to select the entry of the calculated Low Limit and then the up do...

Page 50: ...Cycle Tests Test Locations Hide Disabled Tests Show All Tests Note Bold denotes normal default conditions when shipped from Soft Off Loud factory Beep on Key Hit On Off RS232 Interface 9600 19200 Disa...

Page 51: ...e display NOTE When performing the Get Offset function the test leads HV OUT RETURN should be in their normal position with no device connected for AC Hipot DC Hipot or IR Mode and test leads RETURN G...

Page 52: ...d the right arrow key to advance right through all six digits Note the entered password word all six digits before exiting the last digit The display will now indicate that the Keypad Lockout is On 6...

Page 53: ...ations The Guardian 2500 allows operator selection to show all programmed test locations 1 25 and S1 S9 including those with no test programmed or only those that are enabled 1 Press the UTILITY key a...

Page 54: ...the UTILITY key to exit or move to other Utility functions with the right arrow key 2 4 6 Beep on Key Hit The Guardian 2500 allows operator selection of a key beep on or off 1 Press the UTILITY key an...

Page 55: ...REMOTE HARDWARE NOT DETECTED When the RS 232 is set to Disable with the IEEE 488 interface option present this port is functional refer to paragraph 2 4 8 for available IEEE address selections 2 4 8...

Page 56: ...s from being displayed on the Utility menu It does not change the operation of any of these enhanced features 1 Test Low Limit This selection determines when the low limit is checked If Continuous is...

Page 57: ...hanced Features If Enhanced Features is set for On press right arrow key to select Delay Time If Enhanced Features is set for Off press up down arrow key to select On and press right arrow key to sele...

Page 58: ...s up down arrow keys to select Continuous Voltage on Fail On or Off Press the UTILITY key to exit or move to other Enhanced Features with the right arrow key 4 Continue Steps On Fail This selection de...

Page 59: ...le in IR mode Press the UTILITY key and right arrow key to select Enhanced Features If Enhanced Features is set for On press right arrow key to select Stop Test On Pass If Enhanced Features is set for...

Page 60: ...If Enhanced Features is set for On press right arrow key to select Step And Increment If Enhanced Features is set for Off press up down arrow key to select On and press right arrow key to select Step...

Page 61: ...tinue with Clear All Tests Press down arrow keys to select Clear All Tests No or Yes Press UTILITY key to exit or right arrow key to continue with Clear All Tests if Yes was selected All tests would b...

Page 62: ...elect Tests to View Press up down arrow keys to select Tests To View All Tests 1 25 and S1 S9 Only Single Step 1 25 Only Sequence S1 S9 Press the UTILITY key to exit or move to other Enhanced Features...

Page 63: ...ed the Return input is at chassis ground rather than virtual ground Press the UTILITY key and right arrow key to select Enhanced Features If Enhanced Features is set for On press right arrow key to se...

Page 64: ...ght Arrow key 12 Adaptive DC Ramp This function not available on the Guardian 2510 is intended to optimize measurement speed Off should always be selected for DC Hipot and insulation resistance IR if...

Page 65: ...nging The Guardian 2530 uses seven ranges of sensitivity for the current input to cover its IR range of 10 kohms to 2000 Gohms The lowest resistance which can be measured in any range must be high eno...

Page 66: ...would be used if the DUT IR was exactly equal to the programmed LOW LIMIT for the test in progress 3 Autoranging is being defeated by a non DC component of the input current This may result from oper...

Page 67: ...tup 24 Mode AC Total Current Volt 1 500KV Hi Limit 5 000mA Low Limit Off Arc Limit Off Ramp Time 0sec Hi Limit Ramp Off Low Limit Ramp Off Test Time 1 0sec Fall Time Off Frequency 50Hz Gnd Continuity...

Page 68: ...ed Press STOP to clear GROUND CONTINUITY LIMIT TOO HIGH Handler start ignored Invalid test number High limit changed to be greater than low limit High limit can not be greater than max current High li...

Page 69: ...und fault GFI or overload SHORT Current exceeded power amp limit of 25 mA SHORT Current exceeded limit specified for instrument 20 mAAC 8 mADC SHORT High voltage relay opened because of current surge...

Page 70: ...E HARDWARE NOT DETECTED DAC OVER RANGE EEPROM CALIBRATION DEFAULTS BEING SET EEPROM PERSONALITY DEFAULT BEING SET EEPROM PERSONALITY BEING SET FROM REMOTE EEPROM REMOTE OPTIONS DEFAULTS BEING SET EEPR...

Page 71: ...71 Instrument Error Messages continued Invalid calibration code INVALID DAC SLOPE MEASUREMENT TIMED OUT A D converter failure NVRAM FIRST TIME INIT FAILED NVRAM INIT FROM FLASH Press STOP to continue...

Page 72: ......

Page 73: ...ure below Pass Fail and Under Test relays are rated at 36V 100mA max 1 billion cycles 3 2 Remote I O Figure 3 1 Rear Panel Remote Connectors The Pass and Fail outputs track the front panel PASS and FA...

Page 74: ...s three lines receive data transmit data and signal ground With some controllers additional signals maybe required and are listed in Table 3 1 Refer to Figure 3 3 for null modem cable configuration to...

Page 75: ...ion Pin Function 3 2 6 5 4 Receive data Transmit data Data terminal ready Signal ground Data set ready Connect to 3 2 6 5 4 Transmit data Receive data Data set ready Signal ground Data terminal ready...

Page 76: ...njunction with Pass1 relay closure indicating pass condition Relay remains closed until next test started Fail1 5 Used in conjunction with Fail2 relay closure indicating fail condition Relay remains c...

Page 77: ...sts Value is read when Start is initiated This line must be set 20msec before start and held for 20msec after Data1 10 Binary code 2 to determine test number Data2 11 Binary code 4 to determine test n...

Page 78: ...ndard 488 2 Standard Digital Interface for Programmable Instrumentation Refer to Table 3 3 below for a full tabulation of connections and Table 3 4 for the command set The following functions have bee...

Page 79: ...ace system to a known quiescent state SRQ 10 Service Request Low state a talker or listener signals to the controller need for attention in the midst of the current sequence of events REN 17 Remote En...

Page 80: ...e ARC limit in mA nn nnn OFF TRAmp Set the Ramp time in seconds nnn n OFF RHIGh Set Ramp High limit in mA nn nnn OFF RLOW Set Ramp Low limit in mA nn nnn OFF TDWell Set the Dwell time in seconds nnn n...

Page 81: ...ocations HIDE ALL BUZZer Turn on buzzer on failure LOUD SOFT OFF BEEP Beep when a key is pressed OFF ON ENHanced Show enhanced features on display OFF ON LOWLimit Low Limit is tested continuous or EOT...

Page 82: ...r Off CR Low Lim Ramp tab nn nnnmA or Off CR Dwell Time tab nnn nsec or Off CR Test Time tab nnn nsec or Continuous CR Fall Time tab nnn nsec or Off CR Frequency tab 50Hz or 60Hz CR Gnd Continuity tab...

Page 83: ...y View the utilities programmed If enhanced features is set to On the format of the returned data is as follows Meas Offset tab Off On or Get CR Keypad Lockout tab Off On or On With Test Number CR Tes...

Page 84: ...year mm dd yyyy STOP Stops a measurement MEASure Triggers a measurement of single or sequence test FETCh Fetches the most recent measurement results The character sequence formats are as follows Norma...

Page 85: ...e 1st character of the string is required to determine the status of the tester 0 No status testing idle or testing in progress 1 Failure failure due to short GFI or limit failures 2 No measurement av...

Page 86: ...r value SRE Set the service request enable register value RST Reset the buffer CLS Clear standard event status register OPC Operation complete OPC Is operation complete WAIT Wait until operation is co...

Page 87: ...h IEEE 488 the unit sets SRQ to request permission to send the prompt In multidrop environments poll SRQ for the configured bus address until the MAV bit is set then send a query to retrieve the promp...

Page 88: ...terminated by the NL character together with the EOI line asserted Status Byte Register Decimal Bit Value Use 7 128 Measure completed 6 64 SRQ SPOL Resets 5 32 Summary of Standard Event Status Regist...

Page 89: ...s done in the Standard Event Status Enable Register by issuing an ESE command per paragraph 11 5 1 3 3 5 5 IEEE 488 Sample Program In Quick Basic REM INCLUDE qbdecl bas declarations ADAP GPIB0 G2500 D...

Page 90: ...INPUT TIM IF TIM OR TIM 999 9 THEN GOTO 300 check time is in range configure g2500 IF MODE 1 THEN MODECH AC IF MODE 2 THEN MODECH ACRE IF MODE 3 THEN MODECH ACIM IF MODE 4 THEN MODECH DC conf TEST TES...

Page 91: ...de a Purchase Order Number and credit card information to insure expedient processing although units found to be in warranty will be repaired at no charge For any questions on repair costs or shipment...

Page 92: ...r 0 5 or reading to 3000V for DC and 3000 RMS AC Valhalla 4600 or equivalent suggested Calibration Procedure From the UTILITY key select enhanced features From enhanced features select Perform Calibra...