Quadbeam Technologies MXD75 Quick Start Manual Download Page 1

Fitted Options-    Power Supply: 

  24V 

Output Option Card:      1-Current 2-Relays       Channel1:           Cond  

Channel2:           Cond                    Channel3:            Cond

 

 

 

  Mains 

 

 

    3-Current 0-Relays   

             pH/Redox 

 

        pH/Redox 

 

  pH/Redox

 

 

 

 

 

 

    3-Current 4-Relays   

             ECS   

 

        ECS 

 

 

  ECS

 

 

 

 

 

 

    5-Current 2-Relays   

             DO   

 

        DO 

 

                            DO

 

 

 

 

 

 

    Modbus 4-Relays 

 

             Aux mA IP 

 

        Aux mA IP 

 

  Aux mA IP 

 

 

 

 

 

 

 

 

 

             SS   

 

        SS 

 

 

  SS

MXD75 

Quick Start Guide

Relay Connection Details

MXD75 Instrument Dimensions

74 mm

300 mm

6 mm x 

3 off

8 mm

74 mm

Power Connection Details

CAUTION!

Always remove the main power from the 

system before any alterations to the wiring. 

Ensure that both power lines are isolated. Make 

sure that the power cannot be switched on by 

accident whilst the instrument is being 

connected. For safety reasons an earth 

connection must be made to the earth terminal 

of this instrument.

Local wiring and safety regulations should be 

strictly adhered to when installing this 

instrument. If the installation methods and 

cable types recommended in this guide are 

followed, then the instrument will achieve the 

levels of EMC protection as specified in the 

accompanying CD manual. 

Consult the serial label on the side of the 

instrument for supply voltage requirements.

Power Supply

“Live” In

Power Supply

“Live” Out

(For Daisy Chaining)

Power Supply

“Neutral” In

Power Supply 

“Neutral” Out

(For Daisy Chaining)

Protective Earth

(Must Be Connected)

Earth

L

+

N

-

L

+

N

-

Earth

Earth

Power Connector

85V-265V AC/DC  Power Supply 

Connection Details

Power Supply

“+” In

Power Supply

“+” Out

(For Daisy

Chaining)

Power Supply 

“-” In

Power Supply

“-” Out

(For Daisy 

Chaining)

Protective Earth

(Must Be Connected)

Earth

L

+

N

-

L

+

N

-

Earth

Earth

Power Connector

18V-32V AC/DC  Power Supply 

Connection Details

Relay Connection Details

Available Relays Vary Depending On Instrument Configuration

5-6

NO3 C3 NC3

NO2 C2 NC2

NO1 C1 NC1

NO6

C6

NO5

C5

NO4 C4 NC4

Relay 1

Normally

Open

Contact

Relay 1

Common

Contact

Relay 1

Normally

Closed

Contact

Relay 2

Normally

Open

Contact

Relay 2

Common

Contact

Relay 2

Normally

Closed

Contact

NO1 C1 NC1 NO2 C2 NC2

NC3

NO3 C3

Relay 3

Common

Contact

Relay 3

Normally

Open

Contact

Relay 3

Normally

Closed

Contact

Relay 4

Normally

Open

Contact

Relay 4

Common

Contact

Relay 4

Normally

Closed

Contact

Relay 5

Normally

Open

Contact

Relay 5

Common

Contact

Relay 6

Normally

Closed

Contact

NO4 C4 NC4 C5 NO5 C6 NO6

Relay 6

Normally

Open

Contact

Relays 1 - 3 Connector

Relays 4 - 6 Connector

LTH recommends using No. 

10 x 1¼ inch round head 

screws or similar for 

mounting.

Ensure top screw head is 8 

mm proud.

Care must be taken when 

fitting the unit on uneven 

walls or surfaces. Do not 

over stress the mounting 

lugs.

Over tightening the 

mounting screws could also 

break the lugs.

Drill Hole Dimensions

110 mm

CH3

CH2

CH1

85-265V    15W

_

~

1

4

7

2

5

8

3

6

G

G

G

DIGITAL
INPUTS

SENSOR INPUT 3

SENSOR INPUT 2

SENSOR INPUT 1

L

+

N

-

L

+

N

-

SD CARD

0/4-20 mA

OUTPUT

1

2

A+

-

A

A

B

B G E C

A

+

-

+

-

E

E

R R

A

B

B

A

A

B

B

+

-

+

-

R R

D D

CONDUCTIVITY

PH / REDOX

ECS CONDUCTIVITY

5-6

A+

B+

C+

D+ -

-

-

0/4-20 mA
OUTPUTS

3

2

1

6

5

4

Power Supply

Input

Relays

Current Outputs

/ Modbus RS485

Channel 1

 Input

Channel 2

Input

Channel 3

Input

Digital

Inputs SD Card

 Slot

Earth Studs

Ethernet

Port

(Future

Expansion)

Exact Instrument Configuration Depends

Upon Ordered Specification

MON 1 JUN 2009
09:56

CH1: 517.2 µS/cm TEMP1: 12.3°C
CH2: 9.64 pH
CH3: 23.3%

TEMP2: 25.0°C
TEMP3: 28.0°C

MAIN MENU

DIGITAL INPUTS

CHANNELS

EXIT

4-20mA OUTPUTS

CALIBRATION

SETPOINT / RELAYS

CONFIGURATION

CURSOR

FURTHER MENU 

PAGES BELOW

INSTRUMENT

READINGS

CURRENT

DATE / TIME

UNIT STATUS

FURTHER MENU 

PAGES ABOVE

MON 1 JUN 2009
09:56

CH1: 517.2 µS/cm TEMP1: 12.3°C
CH2: 9.64 pH
CH3: 23.3%

TEMP2: 25.0°C
TEMP3: 28.0°C

SETPOINT 1 SETUP

INPUT SOURCE:

RANGE:

TRIGGER:

HIGH VALUE:

MODE:

CHANNEL:

SENSOR

0 to 9.999 mS/cm

HIGH

7.500mS/cm

ON/OFF

EXIT

CHANNEL 1 (COND)

LOW

BAND

LATCH HIGH

TRIGGER

HIGH

LATCH LOW

CURSOR

ADDITIONAL

OPTIONS BELOW

OPTION POP-UP

MON 1 JUN 2009
09:56

CH1: 517.2 µS/cm TEMP1: 12.3°C
CH2: 9.64 pH
CH3: 23.3%

TEMP2: 25.0°C
TEMP3: 28.0°C

CHANNEL 1 SETUP

UNITS:

CELL CONSTANT:

RANGE:

TEMP INPUT SENSOR:

TEMPERATURE UNITS:

MODE:

SIEMENS(S/cm)

1.000

AUTO

PT1000

°C

EXIT

ON-LINE

SECURITY CODE

CODE

ENTER ACCESS

* * *

0

User Interface Overview

Scrolling Menu Layout

Front Layout

Main Menu Layout

Pop-Up Option Layout

Security Access Pop-Up

To protect the instrument setup 

from unauthorised or accidental 

tampering, a security access code 

system is present. This is 

implemented via the instrument’s 

menu system which operates in two 

modes, “locked” as indicated by a 

padlock     symbol and “unlocked” as 

indicated by a key     symbol. 

Fourth Issue September 2016

For configuration and calibration information please consult the operational manuals 

on the accompanying CD.

The default Access Code is: 1000

Front Screen

Current

Output

Bargraphs

Channels

Setup

Setpoint /

Relays Setup

4-20mA

Outputs Setup

Digital

Inputs Setup

Error

Messages

Live Trends

1, 2 & 3*

*Optional Extra

CAL

SETPOINT STATUS

LEDS

SETPOINT STATUS

LEDS

ALARM STATUS

LEDS

09:56

MON 1 JUN 2009

CH1:

517.2

 

µS/cm

i)

ii)

12.3

°C

A: 12.28mA

CH2:

9.64 

pH

i)

ii)

25.0

°C

-156mV

CH3:

23.3 

%

i)

ii)

28.0

°C

2.15Atm

MENU

CAL

CONTEXT SENSITIVE BUTTON FUNCTIONS

UNIT STATUS

CURRENT TIME

CHANNEL

MESSAGES

CHANNEL SECONDARY 

READING ii 

CURRENT DATE

CHANNEL PRIMARY

READING

CHANNEL ID

CHANNEL SECONDARY

READING i

CHANNEL ALARM

STATUS

TANK 1

TANK 3

CHANNEL

LABEL

A+

C+

D+

E+

F+

-

-

-

0/4-20 mA

OUTPUTS

B+

Current Output Connector

Current 

Output A+

Current

Output B+

B+

C+

D+

E+

F+

Current

Output D+

Current

Output -

(Common)

Current

Output E+

Current

Output C+

Current

Output F+

Current

Output -

(Common)

Current Output Connection Details

A+

-

-

-

Current

Output -

(Common)

Current Output & Modbus Connection Details

-

MODBUS RS485

& 0/4-20 mA OUTPUT

Modbus and Current Output Connector

Current 

Output A+

Current

Output -

(Common)

Modbus A

Conenction

(Inverting)

(Rx-/Tx-) Modbus B

Connection

(Non-Inverting)

(Rx+/Tx+)

Modbus and Current Output

Connection Details

A+

mA

Modbus C

Connection

(0V Reference)

Optional

-

mA

A-

485

B+

485

C

G

485

A+

mA

A-

485

B+

485

C

G

485

mA

Available Current Output and Modbus Connections Vary Depending Upon

Instrument Configuration

Temperature Connection Details

Sensor Input Connector

Link

Link

RTD

2 Wire RTD

Temperature Connection

A

A

B

B

Sensor Input Connector

RTD

3 Wire RTD

Temperature Connection

Link

A

A

B

B

Sensor Input Connector

Link

22K Thermistor

Temperature Connection

Link

Thermistor

A

A

B

B

T

(Only Available On Dissolved Oxygen Input Card)

Sensor Input Connector

RTD

4 Wire RTD

Temperature Connection

A

A

B

B

Digital Input Connection Details

G

1

2

3

4

5

6

7

8

Digital Input Connector

1

4

7

2

5

8

3

6

G

G

G

Digital Input Connection Details

PO Box 1142

Pukekohe

Auckland

2340

New Zealand

Telephone: +64 (0) 9 234 4609

email: [email protected]

Web: www.quadbeam.co.nz

SKT5

SKT6

SKT7

HEADER A

HEADER B

HEADER C

HEADER D OR E

COMMS A

COMMS B

CH3

CH2

CH1

85-250V    15W

_

~

1

4

7

2

5

8

3

6

G

G

G

DIGITAL
INPUTS

SENSOR INPUT 3

SENSOR INPUT 2

SENSOR INPUT 1

L

+

N

-

L

+

N

-

SD CARD

LINK
ACT

0/4-20 mA

OUTPUT

1

2

A+

-

A

A

B

B G E C

A

+

-

+

-

E

E

R R

A

B

B

A

A

B

B

+

-

+

-

R R

D D

CONDUCTIVITY

PH / REDOX

ECS CONDUCTIVITY

5-6

A+

B+

C+

D+ -

-

-

0/4-20 mA
OUTPUTS

3

2

1

6

5

4

F1

F2

F3

F4

F5

09:56

MON 1 JUN 2009

CH1:

517.2

 

µS/cm

i)

ii)

12.3°C

A: 12.28mA

CH2:

9.64 

pH

i)

ii)

25.0°C

-156mV

CH3:

23.3 

%

i)

ii)

28.0°C

2.15Atm

MENU

CAL

242 mm

331 mm

09:56

MON 1 JUN 2009

CH1:

517.2

 

µS/cm

i)

ii)

12.3°C

A: 12.28mA

CH2:

9.64 

pH

i)

ii)

25.0°C

-156mV

CH3:

23.3 

%

i)

ii)

28.0°C

2.15Atm

MENU

CAL