Important
Pausing a run interrupts the sample preparation procedure. Only pause a run in
an emergency.
Important
Pausing a run results in the processed samples being flagged as “unclear”.
8.7.3
Resuming a run
To resume a run, press the
Continue SP
button. If the run was paused, the samples will be flagged
as “unclear”.
8.7.4
Stopping a run
Press the
Stop SP
button to stop the run. All batches currently being processed will be stopped.
Other batches with the status
QUEUED
can be processed in another run after following the
cleanup procedure.
If the run is stopped, all processed samples are flagged with “invalid”. It is not possible to process
these samples further or to resume the run.
Canceled run
After stopping a run or if the run stops due to an error, the “S” button is flashing (see Section
5.1.1.2). Press the “S” button to display the warning or error message. This message gives the
hint to perform the cleanup procedure in the
Maintenance SP
screen.
Important
After successful cleanup, it is necessary to empty all slot positions in the
“Sample” and “Eluate” drawers. New runs can then be defined and started.