QIAsymphony SP/AS Consolidated Operating Guide 12/2017
28
Active logout
To log out, press
Log Out
at the top of the
Sample Preparation
or
Assay
Setup
screen.
If you are logged out, the status bar only displays the date and time.
Automatic logout
After a defined period of user inactivity, the user currently logged in is automatically logged out.
The default setting for this period of user inactivity is 15 minutes. Ask the “Supervisor” to adjust
the time period to suit your needs or to switch it off, if required.
3.3.3
Switching off the QIAsymphony SP/AS
To switch off the QIAsymphony SP/AS instruments, press the power switch at the front of the
QIAsymphony SP in the lower left corner. We recommend switching off the instruments after use.
Important
Do not switch off the instruments during sample preparation or assay setup
unless you need to stop the instruments due to an emergency. You will not be
able to resume the protocol or assay run and the samples cannot be processed
further by the QIAsymphony SP/AS.
Important
The QIAsymphony SP/AS instruments will lose all inventory information when the
instruments are switched off.
Important
After the QIAsymphony SP/AS instruments are switched off, the power switch
flashes a few times. When the power switch stops flashing it is safe to switch the
QIAsymphony SP/AS instruments on again.
Summary of Contents for QIAsymphony SP
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...