QIAsymphony SP/AS Consolidated Operating Guide 12/2017
54
8
Loading QIAsymphony SP Drawers
This section describes how to load and unload the worktable and how to perform inventory scans
to operate the QIAsymphony SP instrument.
Workflow loading of QIAsymphony drawers.
We recommend loading the drawers in the order:
1.
“Waste” drawer
2.
“Eluate” drawer
3.
“Reagents and Consumables” drawer
4.
“Sample” drawer
8.1
Using the software Wizard
The QIAsymphony SP operating software offers a
Wizard
that provides step-
by-step guidance for setting up a run.
The
Wizard
takes you through:
Loading the “Waste” drawer
Loading the “Eluate” drawer
Loading the “Reagents and Consumables” drawer
Loading the “Sample” drawer
Defining a batch/run with or without work lists
Loading internal controls
It is possible to set up a run on the QIAsymphony SP with or without the
Wizard
.
Important
The
Wizard
can only be used for setting up independent sample preparation
runs. It cannot be used to set up an integrated run.
Summary of Contents for QIAsymphony AS
Page 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Page 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Page 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...