QIAcube Connect MDx User Manual 12/2020
97
1.
Select the
Configuration
icon (
).
2.
Press the
Protocols
tab.
Protocols configuration screen
3.
To delete all protocols installed on the instrument, press
Delete All
. It is not possible to delete a
single protocol from the QIAcube Connect MDx.
Note
: After deleting all protocols, a selective upload of protocols will reduce the options during run
setup. This is done by copying only the desired protocols from the backup you created to the
Protocol_Upload
folder on the USB stick.
5.11.3
Saving protocols
Protocols can be downloaded from the instrument to USB flash drive to transfer them to another
instrument or to save them before a software update. Use the USB flash drive provided by QIAGEN.
1.
Connect the USB flash drive that was shipped together with the instrument to the QIAcube
Connect MDx using one of the USB ports at the left of the touchscreen.
2.
Select the
Configuration
icon (
).
3.
Press the
Protocols
tab.
4.
From the
Download Protocols
section, press
Copy To USB
.
Important
: Transferring of DSP protocols to a life science instrument is not allowed. This will
result to loss of the IVD status of the workflow.