QIAcube Connect MDx User Manual 12/2020
88
If a run is stopped, perform the daily maintenance as described in section 6.4, Daily maintenance
procedure, and ensure that no plastic parts are present in the centrifuge before starting the next
run. Additionally, it is recommended to reboot the system before starting the next run.
Note
: If a protocol run is stopped, the run cannot be restarted. Under the
Run Details
, you will find
the step at which the protocol stopped.
5.7
Saving run reports to the USB flash drive
Run reports are saved on the instrument after each run is confirmed by clicking the
Done
button.
The number of run reports on the instrument is limited. If the remaining space on the internal storage
reaches 10% of the total capacity, you are prompted to perform a backup of report files.
Warning on critical remaining disk space
On a regular basis, you can also check the free disk space by clicking on the
Configuration
icon
located on the right side of the
System
tab screen.
Free disk space information