CAUTION
Risk of overheating
[C9]
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the QIAcube Connect MDx.
Slits and openings that ensure the ventilation of the QIAcube Connect MDx
must not be covered.
WARNING
Risk of personal injury and material damage
[W2]
The QIAcube Connect MDx is too heavy to be lifted by one person. To
avoid personal injury or damage to the instrument, do not lift the instrument
alone.
CAUTION
Damage to the instrument
[C6]
Direct sunlight may bleach parts of the instrument and cause damage to plastic
parts.
The QIAcube Connect MDx must be located out of direct sunlight.
4.1.2
Power requirements
The QIAcube Connect MDx operates at: 100–240 V AC, 50/60 Hz, 650 VA. It can be connected
to an uninterruptible power supply (UPS).
Ensure that the voltage rating of the QIAcube Connect MDx is compatible with the AC voltage
available at the installation site. Mains supply voltage fluctuations are not to exceed 10% of nominal
supply voltages.
WARNING
Damage to electronics
[W9]
Before powering ON the instrument, make sure that the correct supply voltage
is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the specifications
indicated in the type plate of the instrument.